Messages
Message
Artifact
●
Asystole
●
Bigeminy
●
Brady
●
Cable off
●
Change telemetry
●
battery
Duplicate TTX
●
ECG error
●
ECG measurements
●
removed
Failure connecting to
●
transmitter
Frequent PVCs
●
Frequent SVCs
●
Gender is not defined
●
HR(ECG) high /
●
HR(ECG) low
464
Location
Explanation
wavef.
Muscle artifact or high or low
●
frequency noise.
al. area,
Physiological alarm.
●
param.,
wavef.
al. area,
Physiological alarm.
●
wavef.
al. area,
Physiological alarm.
●
wavef.
report
●
ECG cable is off.
al. area
The telemetry transmitter's
●
battery has little charge left.
al. area
Multiple patients have been
●
admitted using the same TTX
number.
al. area
The module has a
●
communication error.
al. area
Acquisition module has lost ECG
●
communication.
al.area
Connection to the telemetry
●
transmitter could not be
established.
al. area
Physiological alarm.
●
al. area,
Physiological alarm.
●
wavef.
report
●
ACI-TIPI has been selected on,
but the patient's gender has not
been entered.
al. area
Measurement values are equal
●
to or outside the alarm limits.
CARESCAPE Modular Monitors
What to do
Check electrode contact.
●
Check lead placement.
●
Perform skin preparation.
●
Reposition/replace
●
electrodes.
Request the patient to remain
●
still.
Check the patient status.
●
Check the patient status.
●
Check the patient status.
●
Connect the cable.
●
Change the battery.
●
Contact authorized service
●
personnel.
Check the module.
●
Replace the module if
●
necessary.
Remove the module and then
●
reconnect it.
Replace the module.
●
If the problem persists,
●
contact authorized service
personnel.
Retry connecting.
●
If the problem persists,
●
contact authorized service
personnel.
Check the patient status.
●
Check the patient status.
●
Enter the patient's gender.
●
Check the patient status.
●
Adjust alarm limits if
●
necessary.
2062971-001