Interfacing with peripheral devices
The monitor displays Ppeak, Pplat, Pmean, PEEPtot, PEEPi, PEEP, TVexp, TVinsp,
MVexp, MVspont, Compl, Raw, I:E ratio, and static Compl measurement data.
If there is a compact airway module measuring spirometry in the monitor, it is
used as the measurement source instead of the interfaced device.
Gas data from ventilator, anesthesia machines, and gas analyzers
The monitor displays CO
Isoflurane and Desflurane measurement data.
If there is a compact airway module in the monitor, it is used as the
measurement source instead of the interfaced device.
Laboratory data
The monitor provides Laboratory Data >View and Enter Data menus.
Continuous cardiac output
The interfaced continuous cardiac output measurement provides C.O., blood
temperature, CCO, SVV, GEDI, ELWI and SVR measurements depending on
the interfaced device. The Ext indicates the data is from an external source.
The following conditions apply with the interfaced continuous cardiac output
measurement:
Ventilator/anesthesia machine settings and technical alarms
The settings are displayed in numeric trends.
Technical alarms and settings are displayed and trended even when there is a
gas module connected to the monitor.
452
, Resp Rate, O
2
The C.I. value for interfaced device is computed using the following equation
C.I. = C.O. / BSA.
The C.I. value for interfaced device is invalid if BSA from the monitor is not
available.
The CCI value for interfaced device is computed using the following equation
CCI = CCO / BSA.
The CCI value for interfaced device is invalid if BSA from the monitor is not
available.
The SVR value for the interfaced device shall be computed using the
following equation SVR = ((MAP - CVP) / CCO) * 79.92.
The SVR value for the interfaced device shall be invalid if MAP, CVP or CCO
from the monitor are not available.
The SVRI value for interfaced device shall be computed using the following
equation SVRI = ((MAP - CVP) / CCI) * 79.92.
The SVRI value for the interfaced device shall be invalid if MAP, CVP or CCI
from the monitor are not available.
Tblood value is displayed. The Tblood - T1 temperature alarm limit is
adjustable.
CARESCAPE Modular Monitors
, N
O, Halothane, Desflurane, Enflurane,
2
2
2062971-001