Performance Wst
The procedures in this section test the optical performance of the instrument.
The complete specifications to which the H P E5970A is tested are given in
Appendix C. All tests can be performed without access to the interior of thr
instrument.
Required Test Equipment
The equipment required for the performance test is listed in the table below.
Any equipment which satisfies the critical specifications of the equipment given
in the table, may be substituted for the recommended models.
Instrument o r Accessorries
Quantity Recommended Model
Required Characteristic
Lightwave Multimeter
1
HP 8163A
Optical Head Interface Module
1
HP 81533B
Optical Detector Head
1
HP81620A #C01
Optical Detector Head
1
HI' 81524A #C01
Laser Source 860 nm
1
HP 81561MM
860f 10 nm
short term stability
<f 0.01dB
Laser Source 131011560 nm
1
HP
81664SM
1310% 10nm ( f 2nm)
15605 10nm
short term stability
<f 0.006dB
Optical attenuator 860 nrn
Optical attenuator
attenuation
>
60dB
Return Loss
>
26dB
repeatability
<
3~0.04dB
1
HP 8166A #I01
attenuation
>
60dB
Return Loss
>
46dB
repeatability
<
f
0.01dB
Patchcord HMSlO/HMS10.9160pm, SM
2
HP 81101AC
Patchcord HMSlO/IIMSlO,M)/126pm, MM
2
HP 81601AC
U
connector interface
connector adapter
connector adapter
4
HP 81000AI
1
HP 81000AA
1
HI'
81005AA
Performance Test