Emissions that exceed the levels required by CISPR 11 can occur when the equipment is
connected to a test object.
Korea (KCC) ...................................... Class A Equipment
Class A: Equipment meets requirements for industrial electromagnetic wave equipment and
the seller or user
should take notice of it. This equipment is intended for use in business environments and not
to be used in homes.
USA (FCC) ........................................ 47 CFR 15 subpart B.

Electrical Specifications

Modules
NORMA 6003 .............................. 3 Voltage + 3 Current
NORMA6003+ ........................... 3 Voltage + 3 Current + 1 Motor
NORMA 6004 ............................. 4 Voltage + 4 Current
NORMA 6004+ ........................... 4 Voltage + 4 Current + 1 Motor
Sample Rate .................................... 200 kHz
Data Update Rate ............................ 100 ms, 200 ms, 500 ms, 1 s

Voltage

Range .............................................. 10 V, 100 V, 1000 V
Crest Factor ..................................... CF ≤ 2
Overload Capacity ............................. 10 % overload
Input Impedance .............................. 2 MΩ / 10 pF (Typical)
Temperature Coefficient .................. 0.05 x (Spec)/K @ <18°C or >28°C
Bandwidth ........................................ 1000 V range: 500 kHz
(Industrial Broadcasting & Communication Equipment)
This product is considered an exempt device per clause 15.103.
100 V range: 200 kHz
10 V range: 100 kHz
Power Analyzer
Specifications