Information - Invacare ScanBed 755 User Manual

Nursing bed
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Technical part
Invacare® EC-Høng A/S is certified according to DS/EN ISO 9001 and ISO 13485 which ensu-
res that our customers are always supplied with Invacare® EC-Høng A/S products of uniform
quality.
Throughout the entire production process our materials and products are quality controlled
by the operators.
Furthermore there is a final test when the product is fully
assembled.
The operator carrying out the final test, which comprises
of checking all movable parts, motors and castors, issues a
QA number on the product thereby confirming it´s quality.
If the product does not correspond to the quality demands
of Invacare® EC-Høng A/S, it will be discarded.
If, contrary to our expectations, a problem should arise in connection with the delivered
product, please contact your Invacare® supplier.
3.

Information

Please read the entire technical information section carefully before using or servicing the bed.
All indications of right and left are based on a patient lying on his back in the bed.
Please note that there may be sections in this user manual, which are not relevant for your
bed.
The SB755
Apparatus.
The SB755
to EN 1970.
TM
SB755
electrical system is tested and approved in accordance with EN 60601-1 and
EN 60601-1-2 by UL Demko.
Max. patient weight: 200 kg.
The hand control, control unit and motors are protected according to IP 66. A lock cam must
be used on the control unit - if not, Invacare®EC-Høng cannot guarantee the IP protection.
If the functions of the bed change, please send the bed immediately for a check-up according
to the maintenance chart, chapter 12.
UM_1518190_GB_DK_SE_NO_FI_DE_NL_FR_IT_PT_ver06.indd 12
TM
is CE-marked in accordance with directive 93/42/EEC concerning Medical
TM
is tested and approved by TÜV SÜD Product Service GmbH according
Electromagnetic interference between the bed and other electrical products
can occur. To reduce or eliminate such electromagnetic interference,
increase the distance between the bed and the products or switch them off.
This medical bed can be used together with medical electrical equipment
connected intravascularly or intracardially, provided that following points are
respected:
The medical bed should be equipped with means for potential equalization
connection marked out by a symbol shown in the back of this manual.
Medical electrical equipment should not be fixed on the bed´s metallic
accessories such as side rails, lifting pole, drip rod, bed ends ect.
In addition, the medical electrical equipment power supply cord should be
kept clear of the accessories or any other moving part of the bed.
12
QA XXX
2011-11-08 15:23:49

Advertisement

Table of Contents
loading

Table of Contents