11. Maintenance chart

Service and maintenance of the SB755
who have received the necessary instruction or training
S/N (located on mattress support):
Date:
Circlips, cotter pins and plastic
fixing ring checked.
Screws tightened.
Weldings checked.
Side rail locking and moving system tested.
Castor fittings tightened.
Castor brakes checked.
Height adjustment motor checked.
Backrest motor checked.
Thigh section motor checked.
Leg section motor checked.
Cables correctly suspended and undamaged.
Plugs undamaged.
Damaged coating repaired.
Line side rail straps checked
- must not be frayed or cracked.
NB! Line side rails´ gliding system must not
be lubricated with oil - otherwise the bars will
move sluggishly.
Accessories checked.
A service contract can be made in the countries, where Invacare® has its own sales company.
In certain countries Invacare® offers courses in service and maintenance of the SB755
Spare parts lists and additional user manuals are available from the Invacare® homepage.
The motor cables on the mattress support can be replaced by pushing up the lock with a
screwdriver, and pulling out the cable.