User's Manual of Beyond 300/301 Delivery System

5.7 Junction box

Located in the junction box, the Utility Center comes factory preset at 40psi
for Water pressure and 80 psi for Air pressure. All regulator adjustments
should be made with the Master On/Off toggle in the On position.
1. To increase Air and/or Water Pressure:
To increase pressure,pull out on the adjustment cap then rotate clockwise to
the desired pressure.Push in on the adjustment cap to lock in the adjustment.
2. To Decrease Air and/or Water Pressure:
To decrease pressure, pull out on the adjustment cap then rotate
counterclockwise to the desired pressure. Push in on the adjustment cap to
lock in the adjustment.
NOTE:When decreasing pressure, pressure must be relieved from the
system. This may be achieved pressing the syringe buttons to attain an
accurate measurement from the gauges.
18
P/N:8023488