Operations; Water Supply System - ADS Beyond 300 User Manual

Delivery system
Table of Contents

Advertisement

User's Manual of Beyond 300/301 Delivery System

5 Operations

5.1 Water supply system

A 1.5L quick-disconnect water bottle mounted on the front rotary arm ,provide
simple and convenient solution to control water quality.
1. Air Q/D: This air output provides 80 psi
output to run accessories for the user.
Accepts a 3/8in. male quick-disconnect
fitting.
2. Water Q/D: This water output provides 40
psi output to run accessories for the user.
Accepts a 1/4in. male quick-disconnect
fitting.
3. Flow Control: This needle vale adjusts the
water output from the Water Q/D. Rotate
counterclockwise to increase output and
rotate clockwise to decrease output.
4. City/Bottle Selector: This routing valve
controls the water systerm input. With the
toggle up in the "City"position, water enters
the dental unit from the office plumbing.
With the toggle down in the"Bottle"position,
water enters the dental unit from self-
contained water system.
CAUTION: Only water bottles supplied by the manufacturer can be
used .Do not use soft drink bottles which are thin walled and may rupture
when under pressure.
WARNING: Do not attempt to adjust the water bottle pressure. Bottle
pressure is factory pre-set 40psi.Pressurizing the water bottle over 40 psi
may cause the bottle to rupture.