Assigned Protection Factor; He Filter, Prefilter, Spark Arrestor Service Life; Tr-341N Andtr-344N Battery Charger Display - 3M Versaflo TR-300 Series User Instructions

Powered air purifying respirator papr assembly
Hide thumbs Also See for Versaflo TR-300 Series:
Table of Contents

Advertisement

Battery Pack
Chemistry
Run Time:
Standard – TR-330
High capacity – TR-332
Time to recharge:
Standard – TR-330
High capacity – TR-332
Internal charging temperature
Motor/blower Alarms:
Low air flow
Low battery pack voltage
System Alarm
Intrinsic Safety
Latex
For additional specifications refer to the 3M™ Versaflo™ Powered Air Purifying Respirator TR-300 Technical
Specifications sheet.

Assigned Protection Factor

Refer to the User Instructions for the specific head gear to be used to determine the assigned protection factor
(APF) for the TR-300 PAPR system. Consult 3M Technical Data Bulletin #175 for additional information on APFs and
supporting test data.

HE Filter, Prefilter, Spark Arrestor Service Life

The HE filter must be changed if the low airflow alarm is activated, the air flow as shown by the air flow indicator falls
below 6 CFM or the HE filter becomes damaged or wet. The TR-3600 should be replaced when it is dirty or damaged
and whenever the HE filter is replaced. The TR-362 should be replaced whenever it is damaged.
An excessively loaded HE filter, prefilter or spark arrestor/prefilter may also reduce battery pack run time. If run time is
shorter than expected replacing the filters may resolve the problem.

TR-341N andTR-344N Battery Charger Display

Orange LED
On – steady
On – steady
Off
On – Quick Flash
On – Slow Flash
Slow flash – 1 flash per second; Quick flash – 2 per second
sales@bestmaterials.com
Distributed by Best Materials® LLC
Lithium-ion
Approximately 4-6 hours*
Approximately 8 -12 hours*
* Estimated system run time based on testing with a new battery and a
new clean filter at 68°F (20°C). Actual system run time may be extended or
shortened depending on system configuration and environment.
< 3.0 hours
< 3.5 hours
Optimal: 60°F (20°C) to 95°F (35°C)
Range: 32°F (0°C) to 104°F (40°C)
NOTE: Consult a transportation specialist for any requirements or
limitations prior to transporting lithium ion battery packs.
Activates when airflow falls below 6 CFM (170 lpm) for greater than
approximately 30 seconds. If alarm condition continues (airflow remains
below 6 CFM) for approximately 15 minutes, the TR-300 system will
automatically shut down.
Activates when approximately 10-15 minutes of power remains. Power down
the motor/blower and replace the battery pack to reset alarm. This alarm
will also activate if the battery pack temperature reaches 130°F (55°C). See
Troubleshooting section of these User Instructions.
Audible alarms – 85 dBA at 4 inches (10 cm)
The TR-300 assembly is not an intrinsically safe system.
The TR-300 assembly does not contain dry or natural rubber latex
Green LED
Battery Pack Charge Status
Off
Quick charge mode
Less than 90% charge
On - Slow Flash
Trickle charge mode
90% charge
On – Steady
Fully charged
On – Quick Flash
Error - See Trouble Shooting Section
Off
Battery pack internal temperature too hot or cold
|