CALIBRATION FUNCTION

Occasionally the scale's accuracy should be verified by measuring a known weight
of the scale's approximate capacity. Tare the scale by pressing "ZERO/
the weight on the platform and note the reading. Calibrate the scale if necessary.

CALIBRATION PROCEDURE

You must use a 500g calibrated weight ONLY to properly recalibrate the scale.
1. While the scale is off, press and hold the "ON/OFF" button and then press the
"UNIT" button 3 times. Release all buttons. A number between 40000 and
180000 will display.
2. Press the "ZERO/
3. Place a 500g calibrated weight on the scale platform.
4. Press the "ZERO/
5. Remove the weight. Calibration is complete.
Note: If the calibration is unsuccessful, the display will show a number between
4000 and 70000 instead of 500.0 g. If this occurs, repeat the steps above.
" button. The display will show zero.
" button. The display shows "CAl" then 500.0 g.