Technical data

8.2 Construction

8.1.3

Measuring range

The measuring range refers to the temperature limits in which the thermometer can be used
practically for measuring purposes. Depending on the loads at the place of use and the
required accuracies, the actual measuring range may decrease.
Note
Measuring ranges
The application or possible operating temperatures depend on the configuration of the
temperature sensor.
8.2
Construction
Table 8- 13
Torque values for M20 accessories
Torques for M20 accessories
Plastic cable gland
Metal cable gland
Adapter M20 to NPT½
SITRANS TS500 with head types A requires a minimum thickness of the wrench flat of
5 mm. The maximal wrench size is SW24.
Table 8- 14
Connection type
M14 thread
M18 thread
G½ thread; M20 threads
½ inch NPT
Table 8- 15
Head type
Metal head
Plastic heads
56
Plastic head
5.0 Nm
N/A
5.0 Nm
Torque values between device extension and protective tube
Torque values between device head and extension
Metal head
5.0 Nm
5.0 Nm
5.0 Nm
Torque value
25 Nm
40 Nm
50 Nm
Hand-tight and one to two complete turns with a
wrench
Torque value
20 Nm
5 Nm
TS100/TS200/TS300/TS500/TSinsert/TSthermowell
Compact Operating Instructions, 01/2018, A5E03920348-AE
Cable diameter
6 ... 12 mm
5 ... 14 mm
N/A