Aligning Two Precision
Flanges
Figure B-2
Figure B-3
1. If necessary, place slip pins in the top and bottom holes of one flange
(see Figure B-2 below).
Inserting Slip Rings
2. Using the pins as guides, carefully align the flanges, and insert screws in
two opposite corner holes (see Figure B-3 below).
Align Flanges
3. Place a lock washer and nut on each screw, and finger tighten.
4. Insert the remaining two screws.
5. Place a lock washer and nut on each screw, and finger tighten.
6. If necessary, remove the alignment pins.
7. Go to ""Tighten Flange Connections" on page 4."
Connecting the Waveguide Properly
HP 85325A Operating and Service Manual B-3