OPTIONAL ACCESSORIES
V.
OPTIONAL ACCESSORIES
1 . Senarmont Compensator
To be inserted into the compensator slot of the intermediate tube in
place of the
1/4,,1.
& tint plate to measure the retardation with the
accuracy of the lambda
(A)
unit. (Fig. 24)
Fig.24
1) Detecting extinction position
Rotate the stage with the specimen under the crossed Nicols to find
out the direction where the part of the specimen to be measured
appears darkest.
2) Detecting subtraction position
Rotate the stage 45° to bring it from the extinction position to the
diagonal position. Insert the
1/4,1
& tint plate into the optical path,
and confirm that the interference color of the part of the specimen to
be measured changes toward the lower order side. If the color
changes toward the higher order side, rotate the stage another 90°.
3) Measurement
Place the GI F filter on the filter receptacle, and replace the
1/4,1
&
tint plate with the compensator.
Rotate the analyzer so that the part of the specimen to be measured
becomes darkest.
Let the angle of the above analyzer rotation be theta (8) then the
retardation R (nm) will be obtained as follows:
8
R
=
180,,1.
Where A: wave length of the light used for the measurement
When the GIF filter is used: ,,1.=546nm
30