Chapter 7: Site Preparation – Compressed Air and Vacuum Requirements: Compressed Air Requirements
Compressed Air
Requirements
© Agilent Technologies 1989–2002
This section contains:
I
Connecting Air to the
Testhead, 7-2
I
Compressed Air
Specifications, 7-2
I
Air
Quality, 7-3
The Agilent 3070 / 79000 family uses compressed air to
pressurize air cylinders that actuate fixture pull-down
towers. Pull-down towers are used to pull and hold a
fixture down onto the testhead, enabling contact
Table 7-1
Compressed air specifications for 3070 / 79000 systems
Description
Maximum Pressure for Testhead Only
Minimum Pressure for Testhead with Performance Port
Minimum Pressure for Testhead Only
Minimum Flow Rate for Testhead Only
Minimum Flow Rate for Testhead with Performance Port
Relative Humidity allowed
1. l/s is liters per second. STP is standard temperature and pressure. SCFM is standard cubic feet per minute. See the
pressed Air and Vacuum Primer

Connecting Air to the Testhead

Install compressed air lines using rigid line to a point
near the testhead. An air cutoff valve is recommended
Agilent 3070 / 79000 Site Preparation
between the pins on the fixture and the spring-loaded
interface pins in the testhead.

Compressed Air Specifications

Table 7-1
the 3070 / 79000 family systems. The air flow rates
when using Performance Port are the maximum that
could be needed. Some installations might require lower
volumes.
Measurement
1035 kPa (150 psi)
690 kPa (100 psi)
480 kPa (70 psi)
0.66 l/s at STP (1.4 SCFM)
7.1 l/s at STP (15 SCFM)
70% for compressed air (150 psi) at 25 degrees C
50% for compressed air (150 psi) at 40 degrees C
Dew point must be not more than 5 degrees C
on page 7-19.
for situations when the air line must be disconnected.
When the system is installed, connect the air to the
testhead with a flexible line that has a Hansen Series
shows the compressed air specifications for
1
Com-
7-2