Programmer's Quick Reference
Commands and Queries
Command
Query
:SEQuence:SETup
:SEQuence:SETup?
{MASK | STEP}, <number>,
{MASK | STEP}, <number>
<header> <setup_string>
:SEQuence:STEP
:SEQuence:STEP?
<number>
n/a
:SEQuence:TEST?
*SRE
*SRE?
<mask>
n/a
:STATus?
<display>
n/a
*STB?
8–14
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Options and Query Returns
MASK ::= an individual mask sent to the setup string.
STEP ::= an individual step sent to the setup string.
<number> ::= the mask number or step number sent to the setup string.
<header> ::= the type of setup to be sent or returned:
For individual masks, ::= #800001000.
For individual steps, ::= #800000244.
For whole sequences, ::= #800064122.
<setup_string> ::= the setup string to be sent:
For individual masks, := 1000-byte string.
For individual steps, ::= 244-byte string.
For whole sequences, ::= 64122-byte string.
<number> ::= an integer from 1 to 100 in NR1 format
<result> ::= an integer from 0 to 3 in NR1 format:
0 = pass
1 = fail minimum limit line
2 = fail maximum limit line
3 = fail both minimum and maximum limit lines
<mask> ::= sum of all bits that are set, 0,...,255; an integer in NR1
format. <mask> ::= following values:
Bit
Weight
Enables
7
128
Not Used
6
64
RQS - Request Service
5
32
ESB - Event Status Bit
4
16
MAV - Message Available
3
8
Not used
2
4
Not used
1
2
Not used
0
1
Not used
{ON | OFF}
<display> ::=
{CHANnel <n> | PMEMory{1 | 2}} for the 54600/01/02/03/15/16
{CHANnel <n> | PMEMory{1 | 2} | EXTernal} for the 54610
<n> ::=
1 or 2; an integer in NR1 format for the 54600/03/10/15/16
1, 2, 3, or 4; an integer in NR1 format for the 54601/02
<value> ::= 0,...,255; an integer in NR1 format, as shown in the following:
Bit
Weight
Name
Condition
7
128
----
NOT USED
6
64
RQS/MS
0 = instrument has no reason for service
1= instrument is requesting service
5
32
ESB
0 = no event status conditions occurred
1 = enabled event status condition occurred
4
16
MAV
0 = no output messages are ready
1 = an output message is ready
3
8
----
0 = not used
2
4
----
0 = not used
1
2
----
0 = not used
0
1
----
0 = not used