2-3. Measurement Results
1
Patient information The patient information that was entered before measurement.
Heart Rate/
2
Pulse Rate
3
Electrocardiogram
4
Phonocardiogram
5
%MAP
6
AI
7
UT
Blood pressure
8
value
Blood pressure left-
9
right difference
Measured value
10
reliability
11
PWV
12
ABI
13
Graph 1
14
Facility name
15
PVR waveform
Pulsatile variation
16
graph
17
Dr. Support
56
Shows the measured heart rate. If "ECG" was set to "OFF" in the "SELECT
MEASUREMENT SENSOR" settings when the patient's information was
entered, shows the pulse rate. (refer to page 34)
The ECG waveform. If "ECG" was set to "OFF" in the "SELECT
MEASUREMENT SENSOR" settings when the patient's information was
entered, this will not appear (refer to page 34).
The PCG waveform. If "PCG" was set to "OFF" in the "SELECT
MEASUREMENT SENSOR" settings when the patient's information was
entered, this will not appear (refer to page 34).
This value is one of the pulse waveform indices that are calculated from the
blood pressure values.
Refer to "5-10. Explanation of Technical Terms" (page 143)
This indicates the percentage of the pressure wave effected in the carotid artery
wave that is reflected to form the reflected wave.
Refer to "5-10. Explanation of Technical Terms" (page 143); When "PCG" is
"OFF", AI is not displayed.
Time from the start of the pulse wave to its peak. However, for measurement in
the carotid artery, the time from the start of the pulse wave to the peak of the
effected wave or the start point of the reflected wave.
Refer to "5-10. Explanation of Technical Terms" (page 143)
The blood pressure values for the left and right brachium and left and right ankle.
"- - -" is printed if measurement was not possible.
When the difference in the blood pressure values for the right and left brachia is
greater than 16 mmHg, the maximum blood pressure of the brachium with the
lower blood pressure is shaded.
If a measured value appears in parentheses, the accuracy of that value was low
for some reason and it is indicated only for reference.
baPWV: The PWV value calculated from the interval from the start of the
brachial pulse wave to the start of the ankle pulse wave.
Refer to "5-10. Explanation of Technical Terms" (page 143)
Shows the right and left ABI values.
Refer to "5-10. Explanation of Technical Terms" (page 143)
The indices are shown on a graph. Details are explained from page 58.
This is shown if a facility name was entered (refer to page 86).
The obtained pulse wave. Because the printed amplitude is calibrated from the
measured blood pressure values, the printed amplitude may be different than
the amplitude shown on the screen.
This graph shows the pulsatile variation
obtained from each cuff.
a) Synchronization line: Not printed if synchro
measurement was not possible.
b) Inflation upper limit: When the inflation upper
limit is set to other than "AUTO," the setting is
printed.
c) Measurement accuracy: When "Estimate" or
"First Measurement" is displayed, accuracy may be poor.
d)
: Shows the maximum blood pressure value. Not printed if measurement
was not possible.
e) Level meter: Shows the size of the pulse wave.
f) Frame: When constriction of the brachium or ankle is suspected, the frame is
printed in bold.
A Dr. Support is displayed based on the examination results. The content varies
depending on the level setting of "Dr. Support" in the "Print Default Settings" (page 83).
Limit
L-Ank.