6.3 Transport

Before moving the equipment, always
check that it is stable and that the working
area is free from obstacles, such as
objects, animals or people.
The equipment must be carried by two
people having a suitable physical profile
to be able to lift it (see TECHNICAL DATA).
During transport use protective gloves, pull out both side
handles (11), get hold of them firmly and lift the equipment
being careful to keep it balanced all the time.
6.4

Installation

6.4.1 Positioning

The equipment must be installed in premises classified under
group no. 1, pursuant to CEI 64-8 standards, section 710.
It must be positioned on a level ground, with no inclination.
Your working area must be wide enough to be able to carry out all the
necessary manoeuvres both while working and while carrying out routine
maintenance and cleaning (see technical data).
The wheels make small movements easy. However, for the positioning of the
equipment, the surrounding area must have a diameter of at least 0.9 m (2,95
ft).
Do not forget to take into account the length of the power cable (2.5 m/8.19 ft) and of the
handpiece cables (2 m/6.5 ft).
Do not place the equipment near spraying devices or sources of humidity,
such as saunas, showers, hydromassage tubs, etc.
Do not expose the equipment to direct sunlight or major sources of heat.
Do not place the machine near magnetic devices (cash cards, credit cards,
badges, mobile phones).
Once the equipment is in place, lock the front wheels by using the relevant brake system.