Correction Of Cross-Interference (Function 83) - Siemens FIDAMAT 6 Operating Instructions Manual

Gas analyzer for determination of total hydrocarbon content
Table of Contents

Advertisement

6.6.14

Correction of cross-interference (function 83)

Correction of cross-interference is not usually necessary with the FIDAMAT 6. If you still
want a cross-interference correction in particular cases, it is first necessary to distinguish
what type of interference exists.
Press the first softkey. The following function screen appears on the display.
Figure 6-57
Press the first softkey, and select the type of interference. The following distinctions are
possible:
● No interference correction
● Interference correction with const. influence of interfering gas
● Interference correction with variable influence on an. input
● Interference correction with variable influence via ELAN
The cross-interference correction is deactivated for the duration of a calibration process
(zero or span). It becomes active again after calibration is complete and the device has
returned to the "Measuring" operating state.
Interference correction with const. influence of interfering gas
Figure 6-58
FIDAMAT 6
Operating Instructions, 03/2018, A5E00222135-03
Performing a cross-interference correction
Constant influence of interfering gas
Functions
6.6 Configuration