10.5 Dosing function

see section 12

10.6 Outputs

The FM100 flowmeter provides a total of 2 outputs that are freely configurable. The
configuration of the outputs (output 1 and output 2) is done via a wizard function.
The wizard function guides the user step by step through all necessary settings.
Steps:
Select output
Selection of the source or the measurement variable to be output (Flow,
Volume, Temperature, Part volume)
Selection of an output type (4-20 mA, 0-20 mA, 0-10 V, 2-10 V, alarm, pulse,
frequency output,IO-Link, control input)
Setting the output (scaling, thresholds)
Save the configuration
page 26
Sub-
Sub-
Description
para-
para-
meter
meter
level 2
level 3
Sets the flow unit
Sets the value for the
low flow cutoff
Sets the counting
mode
Sets the volume unit
Sets the temperature
unit
Sets the counting
mode
Sets the volume unit
Sets the counter
value to "0"
Value range/
Default
value list
value
LPM
L/m, L/h, m3/h,
galUS/m,
galUS/h,
L/m
galUK/m,
galUK/h, L/s,
mL/s, User
0 ≤ Value ≤
Range start
Range start
absolute
ml, L, m3,
galUS, galUK,
L
User
°C, °F, User
°C
absolute
ml, L, m3,
galUS, galUK,
L
User
A5E48762204-AA
Default
value
GPM
galUs/
m
L
°C
galUs