Endress+Hauser Levelflex FMP55 Technical Information

Guided wave radar
Hide thumbs Also See for Levelflex FMP55:
Table of Contents

Advertisement

Quick Links

TI01003F/00/EN/22.18
71394761
2018-04-12
Products
Technical Information
Levelflex FMP55
Guided wave radar
Interface measurement in liquids
Application
• Rod, rope or coax probe
• Process connection: Flange
• Process temperature: –50 to +200 °C (–58 to +392 °F)
• Process pressure: –1 to +40 bar (–14.5 to +580 psi)
• Maximum measuring range: Rod 4 m (13 ft); rope 10 m (33 ft); coax 6 m (20 ft)
• Accuracy: ±2 mm (±0.08 in)
• International explosion protection certificates; marine approval; EN10204-3.1
• Linearity protocol (3-point, 5-point)
Your benefits
• Reliable measurement even for changing product and process conditions
• HistoROM data management for easy commissioning, maintenance and
diagnostics
• Highest reliability due to Multi-Echo Tracking
• Hardware and software developed according to IEC 61508 (up to SIL3)
• Seamless integration into control or asset management systems
• Intuitive user interface in national languages
• Bluetooth® wireless technology for commissioning, operation and maintenance via
free iOS / Android app SmartBlue
• Easy proof test for SIL
• Heartbeat Technology™
Solutions
Services

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Levelflex FMP55

  • Page 1 Products Solutions Services TI01003F/00/EN/22.18 71394761 2018-04-12 Technical Information Levelflex FMP55 Guided wave radar Interface measurement in liquids Application • Rod, rope or coax probe • Process connection: Flange • Process temperature: –50 to +200 °C (–58 to +392 °F) • Process pressure: –1 to +40 bar (–14.5 to +580 psi) •...
  • Page 2: Table Of Contents

    Levelflex FMP55 Table of contents Important document information ....4 Process ........53 Symbols .
  • Page 3 Levelflex FMP55 Application Packages ..... . . 86 Heartbeat Diagnostics ......
  • Page 4: Important Document Information

    Levelflex FMP55 Important document information Symbols Safety symbols Symbol Meaning DANGER! This symbol alerts you to a dangerous situation. Failure to avoid this situation will DANGER result in serious or fatal injury. WARNING! This symbol alerts you to a dangerous situation. Failure to avoid this situation can WARNING result in serious or fatal injury.
  • Page 5 Levelflex FMP55 Symbols in graphics Symbol Meaning 1, 2, 3 ... Item numbers … Series of steps A, B, C, ... Views A-A, B-B, C-C, ... Sections Hazardous area Indicates a hazardous area. Safe area (non-hazardous area) Indicates the non-hazardous area.
  • Page 6: Terms And Abbreviations

    Time of Flight FieldCare Scalable software tool for device configuration and integrated plant asset management solutions DeviceCare Universal configuration software for Endress+Hauser HART, PROFIBUS, FOUNDATION Fieldbus and Ethernet field devices Device Type Manager Device Description for HART communication protocol ε...
  • Page 7: Registered Trademarks

    Registered trademark of the FieldComm Group, Austin, Texas, USA Bluetooth® The Bluetooth® word mark and logos are registered trademarks owned by the Bluetooth SIG, Inc. and any use of such marks by Endress+Hauser is under license. Other trademarks and trade names are those of their respective owners. Apple®...
  • Page 8: Function And System Design

    Levelflex FMP55 Function and system design Measuring principle Basic principles The Levelflex is a "downward-looking" measuring system that functions according to the ToF method (ToF = Time of Flight). The distance from the reference point to the product surface is measured.
  • Page 9 Levelflex FMP55 Dielectric constant The dielectric constant (DC) of the medium has a direct impact on the degree of reflection of the highfrequency pulses. In the case of large DC values, such as for water or ammonia, there is strong pulse reflection while, with low DC values, such as for hydrocarbons, weak pulse reflection is experienced.
  • Page 10 When using the capacitive measurement of FMP55: • Conductivity of the upper medium: < 1 μS/cm • Conductivity of the lower medium: > 100 μS/cm For FMP55: Under certain conditions measurement is possible even with a changing DC. For details please contact your Endress+Hauser representative. Endress+Hauser...
  • Page 11 • Hardware and software developed according to SIL IEC 61508 • Genuine, direct interface measurement Procurement • Endress+Hauser being the world market leader in level measurement guarantees asset protection • Worldwide support and service Installation • Special tools are not required •...
  • Page 12: Measuring System

    7 m (23 ft) can be ordered. If required, rod and rope probes can be replaced. They are secured with Nord-Lock washers or a thread coating. For further information on service and spare parts please contact the Endress+Hauser service. Endress+Hauser...
  • Page 13: Input

    Levelflex FMP55 Input Measured variable The measured variable is the distance between the reference point and the product surface. Subject to the empty distance entered "E" the level is calculated. Alternatively, the level can be converted into other variables (volume, mass) by means of linearization (32 points).
  • Page 14: Blocking Distance

    Levelflex FMP55 Blocking distance The upper blocking distance (= UB) is the minimum distance from the reference point of the measurement (mounting flange) to the maximum level. 100% A0011279  3 Definition of blocking distance and safety distance Reference point of measurement...
  • Page 15: Output

    Levelflex FMP55 Output Output signal HART Signal coding FSK ±0.5 mA over currency signal Data transmission rate 1200 Baud Galvanic isolation Bluetooth® wireless technology Device version Ordering feature 610 "Accessory mounted", option NF "Bluetooth" Operation / configuration By the SmartBlue app.
  • Page 16: Signal On Alarm

    Levelflex FMP55 Switch output Function Open collector switching output Switching behavior Binary (conductive or non-conductive), switches when the programmable switch point is reached Failure mode non-conductive Electrical connection U = 16 to 35 V , I = 0 to 40 mA...
  • Page 17: Protocol-Specific Data

    Levelflex FMP55 Protocol-specific data HART Manufacturer ID 17 (0x11) Device type ID 0x1122 HART specification Device description files (DTM, Information and files at: • www.endress.com • www.fieldcommgroup.org HART load Min. 250 Ω HART device variables The measured values can be freely assigned to the device variables.
  • Page 18 Levelflex FMP55 PROFIBUS PA Manufacturer ID 17 (0x11) Ident number 0x1558 Profile version 3.02 GSD file Information and files at: • www.endress.com GSD file version • www.profibus.org Output values Analog Input: • Level linearized • Distance • Interface • Interface distance •...
  • Page 19 Levelflex FMP55 FOUNDATION Fieldbus Manufacturer ID 0x452B48 Device type 0x1022 Device Revision 0x01 DD Revision Information and files at: • www.endress.com CFF Revision • www.fieldcommgroup.org Device Tester Version (ITK 6.01 Version) ITK Test Campaign Number IT080500 Link Master (LAS) capable Link Master / Basic Device yes;...
  • Page 20 Levelflex FMP55 Transducer Blocks Block Content Output values Setup Transducer Block Contains all parameters for a standard commissioning • Level or volume procedure (Channel 1) • Distance (Channel 2) Advanced Setup Contains all parameters for a more detailed no output values...
  • Page 21 Levelflex FMP55 Function Blocks Block Content Number of Number of Execution Functionality permanent instantiable time blocks blocks Resource Block The Resource Block contains all the enhanced data that uniquely identify the field device. It is an electronic version of a nameplate of the device.
  • Page 22: Power Supply

    Levelflex FMP55 Power supply Terminal assignment Terminal assignment 2-wire: 4-20 mA HART A0036498  4 Terminal assignment 2-wire: 4-20 mA HART Without integrated overvoltage protection With integrated overvoltage protection Connection 4-20 mA HART passive: terminals 1 and 2, without integrated overvoltage protection...
  • Page 23 Levelflex FMP55 Terminal assignment 2-wire: 4-20 mA HART, switch output A0036500  6 Terminal assignment 2-wire: 4-20 mA HART, switch output Without integrated overvoltage protection With integrated overvoltage protection Connection 4-20 mA HART passive: terminals 1 and 2, without integrated overvoltage protection...
  • Page 24 Levelflex FMP55 Terminal assignment 2-wire: 4-20 mA HART, 4-20 mA A0036500  8 Terminal assignment 2-wire: 4-20 mA HART, 4-20 mA Without integrated overvoltage protection With integrated overvoltage protection Connection current output 1, 4-20 mA HART passive: terminals 1 and 2, without integrated overvoltage...
  • Page 25 Levelflex FMP55 Terminal assignment 4-wire: 4-20 mA HART (10.4 to 48 V A0036516  10 Terminal assignment 4-wire: 4-20 mA HART (10.4 to 48 V Connection 4-20 mA HART (active): terminals 3 and 4 Connection supply voltage: terminals 1 and 2 Terminal for cable screen Block diagram 4-wire: 4-20 mA HART (10.4 to 48 V...
  • Page 26 Levelflex FMP55 Terminal assignment 4-wire: 4-20 mA HART (90 to 253 V A0036519  12 Terminal assignment 4-wire: 4-20 mA HART (90 to 253 V Connection 4-20 mA HART (active): terminals 3 and 4 Connection supply voltage: terminals 1 and 2...
  • Page 27 Levelflex FMP55 Block diagram 4-wire: 4-20 mA HART (90 to 253 V A0036527  13 Block diagram 4-wire: 4-20 mA HART (90 to 253 V Evaluation unit, e.g. PLC HART communication resistor (≥ 250 Ω); observe maximum load Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem) Analog display device;...
  • Page 28 Levelflex FMP55 Block diagram PROFIBUS PA / FOUNDATION Fieldbus A0036530  15 Block diagram PROFIBUS PA / FOUNDATION Fieldbus Cable screen; observe cable specifications Connection PROFIBUS PA / FOUNDATION Fieldbus Measuring device Switch output (open collector) Endress+Hauser...
  • Page 29 Levelflex FMP55 Connection examples for the switch output For HART devices, the switch output is available as an option. See product structure, feature 20: "Power Supply, Output", option B: "2-wire; 4-20 mA HART, switch output" Devices with PROFIBUS PA and FOUNDATION Fieldbus always have a switch output.
  • Page 30 Levelflex FMP55 HART loop converter HMX50 The dynamic variables of the HART protocol can be converted into individual 4 to 20 mA sections using the HART loop converter HMX50. The variables are assigned to the current output and the measuring ranges of the individual parameters are defined in the HMX50.
  • Page 31: Device Plug Connectors

    Levelflex FMP55 Device plug connectors For the versions with fieldbus plug connector (M12 or 7/8"), the signal line can be connected without opening the housing. Pin assignment of the M12 plug connector Meaning Signal + not connected Signal - Ground A0011175 Pin assignment of the 7/8"...
  • Page 32: Power Supply

    Levelflex FMP55 Power supply An external power supply is required. Various supply units can be ordered from Endress+Hauser. 2-wire, 4-20mA HART, passive 2-wire; 4-20mA HART "Approval" Terminal voltage U at the device Maximum load R, depending on the supply voltage U...
  • Page 33 Levelflex FMP55 2-wire; 4-20mA HART, 4-20mA "Approval" Terminal voltage U at the device Maximum load R, depending on the supply voltage U at the supply unit Channel 1: 17 to 30 V R [ ] U 0 [V] A0034973 Channel 2:...
  • Page 34: Power Consumption

    Levelflex FMP55 4-wire, 4-20mA HART, active "Power supply; Output" Terminal voltage Maximum load R K: 4-wire 90-253VAC; 4-20mA HART 90 to 253 V (50 to 60 Hz), overvoltage 500 Ω category II L: 4-wire 10,4-48VDC; 4-20mA HART 10.4 to 48 V...
  • Page 35: Power Supply Failure

    Levelflex FMP55 FOUNDATION Fieldbus Device basic current 15 mA Failure current FDE (Fault 0 mA Disconnection Electronic) FISCO 17.5 V 550 mA 5.5 W 5 nF 10 μH Power supply failure • Configuration is retained in the HistoROM (EEPROM). • Error messages (incl. value of operated hours counter) are stored.
  • Page 36: Overvoltage Protection

    < 1.5 pF Nominal arrest impulse voltage (8/20 μs) 10 kA External overvoltage protection module HAW562 or HAW569 from Endress+Hauser are suited as external overvoltage protection. For detailed information please refer to the following documents: • HAW562: TI01012K • HAW569: TI01013K...
  • Page 37: Performance Characteristics

    Levelflex FMP55 Performance characteristics Reference operating • Temperature = +24 °C (+75 °F) ±5 °C (±9 °F) conditions • Pressure = 960 mbar abs. (14 psia) ±100 mbar (±1.45 psi) • Humidity = 60 % ±15 % • Reflection factor ≥ 0,8 (water surface for coax probe, metal plate for rod and rope probe with min.
  • Page 38: Resolution

    Levelflex FMP55 80 (3.15) 60 (2.36) DC = 80 40 (1.57) 20 (0.79) -20 (-0.79) -40 (-1.57) -60 (-2.36) -80 (-3.15) A0021482  20 Measuring error at the end-of-probe for rope probes Distance from probe end Measuring error: Sum of non-linearity, non-repeatability and hysteresis If for rope probes the DC value is less than 7, then measurement is not possible in the area of the straining weight (0 to 250 mm from end of probe;...
  • Page 39: Reaction Time

    Levelflex FMP55 Reaction time The reaction time can be parametrized. The following step response times (as per DIN EN IEC 61298-2 / DIN EN IEC 60770-1) are valid if the damping is switched off: Level measurement Probe length Sampling rate Step response time <...
  • Page 40: Mounting

    A0011281  22 Mounting position of Levelflex FMP55 • Rod probes / rope probes: must be mounted in a stilling well or bypass →  45. • Coax probes: can be mounted at an arbitrary distance from the wall of the vessel.
  • Page 41 Levelflex FMP55 Applications with restricted mounting space Mounting with remote sensor The device version with a remote sensor is suited for applications with restricted mounting space. In this case the electronics housing is mounted at a separate position from which it is easier accessible.
  • Page 42 Levelflex FMP55 Notes on the mechanical load of the probe Tensile load limit of rope probes Sensor Feature 060 Probe Tensile load limit [kN] FMP55 NA, ND Rope 4mm (1/6") PFA>316 Bending strength of rod probes Sensor Feature 060 Probe...
  • Page 43 Levelflex FMP55 Mounting cladded flanges • Use flange screws according to the number of flange holes. • Tighten the screws with the required torque (see table). • Retighten the screws after 24 hours or after the first temperature cycle. • Depending on process pressure and process temperature check and retighten the screws at regular intervals.
  • Page 44 Levelflex FMP55 Securing the probe Securing coax probes For WHG approvals: For probe lengths ≥ 3 m (10 ft) a support is required. A0012608 Coax probes can be supported at any point of the outer tube. Endress+Hauser...
  • Page 45 Levelflex FMP55 Special mounting conditions Bypasses and stilling wells In bypass and stilling well applications it is recommended to use a centering disks or stars. A0014129 Mounting in a stilling well Mounting in a bypass Minimum distance between end of probe and lower edge of the bypass; see table below...
  • Page 46 With heat insulated tanks the bypass should also be insulated in order to prevent condensate formation. For information on bypass solutions from Endress+Hauser please contact your Endress+Hauser sales representative. Endress+Hauser...
  • Page 47 Levelflex FMP55 Underground tanks A0014142 Use a coax probe for nozzles with large diameters in order to avoid reflections at the nozzle wall. Non-metallic vessels When mounting Levelflex in a non-metallic vessel, use a coax probe. Endress+Hauser...
  • Page 48 Levelflex FMP55 Vessels with heat insulation If process temperatures are high, the device must be included in normal tank insulation to prevent the electronics heating up as a result of heat radiation or convection. The insulation may not exceed beyond the points labeled "MAX" in the drawings.
  • Page 49: Operating Conditions: Environment

    Levelflex FMP55 Operating conditions: Environment Ambient temperature range Measuring device –40 to +80 °C (–40 to +176 °F) Local display –20 to +70 °C (–4 to +158 °F), the readability of the display may be impaired at temperatures outside the temperature range.
  • Page 50 Levelflex FMP55 With a temperature (T ) at the process connection the admissible ambient temperature (T ) is reduced according to the following diagram (temperature derating): Temperature derating for FMP55 [°C] ([°F])T a +80(+176) GT20: +59(+138) +56(+133) GT18: +33(+91) 4...20 mA HART...
  • Page 51 Levelflex FMP55 GT18 = stainless steel housing A = 1 current output = ambient temperature GT19 = plastic housing C = 2 current outputs = temperature at the process connection GT20 = aluminum housing = PROFIBUS PA K, L = 4-wire For PROFIBUS PA and FOUNDATION Fieldbus the temperature derating depends on the usage of the switch output.
  • Page 52: Storage Temperature

    Levelflex FMP55 Storage temperature –40 to +80 °C (–40 to +176 °F) Climate class DIN EN 60068-2-38 (test Z/AD) Altitude according to • Generally up to 2 000 m (6 600 ft) above MSL. • Above 2 000 m (6 600 ft) if the following conditions are met: IEC61010-1 Ed.3...
  • Page 53: Process

    Levelflex FMP55 Process Process temperature range The maximum permitted temperature at the process connection is determined by the O-ring version ordered: Device O-ring material Process temperature FMP55 — –50 to +200 °C (–58 to +392 °F); completely coated High process temperatures (> 150 °C (302 °F)) may possibly accelerate diffusion of the process medium through the probe coating, which may reduce the operating time.
  • Page 54: Mechanical Construction

    Levelflex FMP55 Mechanical construction Dimensions Dimensions of the electronics housing 98 (3.86)* 78 (3.07) 90 (3.54) A0011666  24 Housing GT18 (316L); Dimensions in mm (in) *for devices with integrated overvoltage protection. 99 (3.9)* 78 (3.07) 90 (3.54) A0011346  25 Housing GT19 (Plastics PBT);...
  • Page 55 Levelflex FMP55 Dimensions of the mounting bracket ø42...60 (1.65...2.36) 127...140 122 (4.8) (5...5.51) 161 (6.34) 162...175 (6.38...6.89) A0014793  27 Mounting bracket for the electronics housing. Unit of measurement mm (in) Wall mounting Pipe mounting For the "Sensor remote" device version (see feature 060 of the product structure), the mounting bracket is part of the delivery.
  • Page 56 Levelflex FMP55 FMP55: Dimensions of process connection and probe ø16 (0.63) 2” ANSI B16.5 ø16 (0.63) ø101 (3.98) DN50 EN1092 10K50 JIS B2220 3” ANSI B16.5 ø16 (0.63) DN80 EN1092 ø131 (5.16) 10K80 JIS B2220 4” ANSI B16.5 ø4 (0.16) DN100 EN1092 ø156 (6.14)
  • Page 57: Tolerance Of Probe Length

    Levelflex FMP55 Tolerance of probe length Rod and coax probes Over [m (ft)] — 1 (3,3) 3 (9,8) 6 (20) Up to [m (ft)] 1 (3,3) 3 (9,8) 6 (20) — Admissible tolerance [mm -5 (-0,2) -10 (-0,39) -20 (-0,79)
  • Page 58: Weight

    Levelflex FMP55 Weight Housing Part Weight Housing GT18 - stainless steel approx. 4.5 kg Housing GT19 - plastic approx. 1.2 kg Housing GT20 - aluminium approx. 1.9 kg FMP55 Part Weight Part Weight Sensor approx. 1.2 kg + weight of flange Rod probe 16 mm approx.
  • Page 59: Materials: Gt18 Housing (Stainless Steel, Corrosion- Resistant)

    Levelflex FMP55 Materials: GT18 housing (stainless steel, corrosion- resistant) A0036037 No. Part Material Housing CF3M similar to 316L/1.4404 2.1 Cover of the electronics compartment • Cover: CF3M (similar to 316L/1.4404) • Window: glass • Cover seal: NBR • Seal of the window: NBR •...
  • Page 60: Materials: Gt19 Housing (Plastic)

    Levelflex FMP55 Materials: GT19 housing (plastic) A0013788 No. Part Material Housing 2.1 Cover of the electronics compartment • Cover glass: PC • Cover frame: PBT-PC • Cover seal: EPDM • Thread-coating: Graphite-based lubricant varnish 2.2 Cover of the terminal compartment •...
  • Page 61: Materials: Gt20 Housing

    Levelflex FMP55 Materials: GT20 housing (die-cast aluminum, powder- coated) A0036037 Nr. Part Material Housing, RAL 5012 (blue) • Housing: AlSi10Mg(<0,1% Cu) • Coating: Polyester 2.1 Cover of the electronics compartment; RAL 7035 • Cover: AlSi10Mg(<0,1% Cu) (gray) • Window: Glass •...
  • Page 62 Levelflex FMP55 Nr. Part Material Ground terminal • Screw: A2 • Spring washer: A2 • Clamp: 304 (1.4301) • Holder: 304 (1.4301) Adhesive nameplate Plastic For the version with M12 plug the sealing material is Viton. For the version with 7/8" plug, the sealing material is NBR.
  • Page 63: Materials: Process Connection

    Levelflex FMP55 Materials: Process Endress+Hauser supplies DIN/EN flanges made of stainless steel according to AISI 316L (DIN/EN material number 1.4404 or 1.4435). With regard to their temperature stability connection properties, the materials 1.4404 and 1.4435 are grouped under 13E0 in EN 1092-1:2007 Tab.
  • Page 64: Materials: Probe

    Levelflex FMP55 Materials: Probe Levelflex FMP55 Feature 060 "Probe" • CA: rod 16mm • NA: rope 4mm • UA: ...mm, coax Pos. Material • CB: rod 0,63in • ND: rope 1/6" • UB: ...inch, coax 316L (1.4404) Coating 2 mm (0,08 in): PFA (Daikin PFA AP230) Rope: 316 (1.4401)
  • Page 65: Materials: Mounting Bracket

    Levelflex FMP55 Materials: Mounting bracket A0015143 Mounting bracket for version "Sensor remote" Position Part Material Bracket 316L (1.4404) Bracket 316Ti (1.4571) Screw/nuts A4-70 Distance sleeves 316Ti (1.4571) or 316L (1.4404) Half-shells 316L (1.4404) Endress+Hauser...
  • Page 66: Materials: Adapter And Cable For Remote Sensor

    Levelflex FMP55 Materials: Adapter and cable for remote sensor A0021722 Adapter and cable for version "Sensor remote" Position Part Material Cable FRNC Sensor adapter 304 (1.4301) Clamp 316L (1.4404) Screw A4-70 Loop 316 (1.4401) Crimp sleeve Aluminum Nameplate 304 (1.4301)
  • Page 67: Materials: Weather Protection Cover

    Levelflex FMP55 Materials: Weather protection cover A0015473 Part: Material Protection cover: 316L (1.4404) Molded rubber part (4x): EPDM Clamping screw: 316L (1.4404) + carbon fibre Bracket: 316L (1.4404) Ground terminal • Screw: A4 • Spring washer: A4 • Clamp: 316L (1.4404) •...
  • Page 68: Operability

    Levelflex FMP55 Operability Operating concept Operator-oriented menu structure for user-specific tasks • Commissioning • Operation • Diagnostics • Expert level Operating languages • English • Deutsch • Français • Español • Italiano • Nederlands • Portuguesa • Polski • русский язык (Russian) •...
  • Page 69: Local Operation

    Levelflex FMP55 Local operation Operation Pushbuttons Touch Control with Order code for Option C "SD02" Option E "SD03" "Display; Operation" A0036312 A0036313 Display 4-line display 4-line display elements white background lighting; switches to red in event of device error Format for displaying measured variables and status variables can be individually configured Permitted ambient temperature for the display: –20 to +70 °C (–4 to +158 °F)
  • Page 70: Operation Via Bluetooth® Wireless Technology

    Levelflex FMP55 Operation via Bluetooth® Requirements wireless technology A0036790  31 Device with Bluetooth module Electronics housing of the device Bluetooth module This operation option is only available for devices with Bluetooth module. There are the following options: • The device has been ordered with a Bluetooth module: Feature 610 "Accessory Mounted", option NF "Bluetooth"...
  • Page 71: Remote Operation

    Levelflex FMP55 Remote operation Via HART protocol A0036169  33 Options for remote operation via HART protocol PLC (Programmable Logic Controller) Transmitter power supply unit, e.g. RN221N (with communication resistor) Connection for Commubox FXA191, FXA195 and Field Communicator 375, 475 Field Communicator 475 Computer with operating tool (e.g.
  • Page 72 Levelflex FMP55 Via FOUNDATION Fieldbus FF-HSE FF-H1 FF-H1 A0017188  35 FOUNDATION Fieldbus system architecture with associated components FFblue Bluetooth modem Field Xpert SFX350/SFX370 DeviceCare/FieldCare NI-FF interface card Industrial network FF-HSE High Speed Ethernet FF-H1 FOUNDATION Fieldbus-H1 Linking Device FF-HSE/FF-H1...
  • Page 73 Levelflex FMP55 DeviceCare/FieldCare via service interface (CDI) A0032466  36 DeviceCare/FieldCare via service interface (CDI) Service interface (CDI) of the instrument (= Endress+Hauser Common Data Interface) Commubox FXA291 Computer with DeviceCare/FieldCare operating tool Endress+Hauser...
  • Page 74: Integration In Tank Gauging System

    Levelflex FMP55 Integration in tank gauging The Endress+Hauser Tank Side Monitor NRF81 provides integrated communications for sites with multiple tanks, each with one or more sensors on the tank, such as radar, spot or average system temperature, capacitive probe for water detection and/or pressure sensors. Multiple protocols out of the Tank Side Monitor guarantee connectivity to nearly any of the existing industry standard tank gauging protocols.
  • Page 75: Supplycare Inventory Management Software

    Levelflex FMP55 SupplyCare inventory SupplyCare is a web-based operating program for coordinating the flow of material and information along the supply chain. SupplyCare provides a comprehensive overview of the levels of management software geographically distributed tanks and silos, for instance, providing complete transparency over the current inventory situation, regardless of time and location.
  • Page 76 Levelflex FMP55 A0034288  38 Example of inventory management platform with SupplyCare Enterprise SCE30B SupplyCare Enterprise (via Web browser) SupplyCare Enterprise installation SupplyCare Enterprise on mobile devices (via Web browser) Ethernet/WLAN/UMTS Fieldgate FXA42 Power supply 24 V DC Modbus TCP via Ethernet as server/client...
  • Page 77 Levelflex FMP55 Cloud-based application: SupplyCare Hosting SupplyCare Hosting is offered as a hosting service (software as a service). Here, the software is installed within the Endress+Hauser IT infrastructure and made available to the user in the Endress +Hauser portal. A0034289 ...
  • Page 78: Certificates And Approvals

    CE mark in the corresponding EC Declaration of Conformity together with the standards applied. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. The measuring system complies with the substance restrictions of the Restriction on Hazardous RoHS Substances Directive 2011/65/EU (RoHS 2).
  • Page 79: Pressure Equipment With Allowable Pressure ≤ 200 Bar (2 900 Psi)

    Levelflex FMP55 Pressure equipment with Pressure instruments with a flange and threaded boss that do not have a pressurized housing do not fall within the scope of the Pressure Equipment Directive, irrespective of the maximum allowable allowable pressure pressure. ≤ 200 bar (2 900 psi)
  • Page 80: Test, Certificate

    Levelflex FMP55 Test, Certificate Feature 580 Designation Available for "Test, Certificate" 3.1 Material certificate, wetted metallic parts, EN10204-3.1 inspection FMP55 certificate Conformity to NACE MR0175, wetted metallic parts FMP55 3.1 Material certificate, pressurized parts, EN10204-3.1 inspection FMP55 certificate Conformity to NACE MR0103, wetted metallic parts...
  • Page 81: Other Standards And Guidelines

    Levelflex FMP55 Other standards and • EN 60529 Degrees of protection by housing (IP code) guidelines • EN 61010-1 Protection Measures for Electrical Equipment for Measurement, Control, Regulation and Laboratory Procedures. • IEC/EN 61326 "Emission in accordance with Class A requirements". Electromagnetic compatibility (EMC requirements) •...
  • Page 82: Ordering Information

    • Depending on the device: Direct input of measuring point-specific information such as measuring range or operating language • Automatic verification of exclusion criteria • Automatic creation of the order code and its breakdown in PDF or Excel output format • Ability to order directly in the Endress+Hauser Online Shop Endress+Hauser...
  • Page 83: 3-Point Linearity Protocol

    Levelflex FMP55 3-point linearity protocol The following notes must be taken into account if option F3 ("3 point linearity protocol") has been selected in feature 550 ("Calibration"). Depending on the probe the 3 points of the linearity protocol are defined as follows:...
  • Page 84: 5-Point Linearity Protocol

    Levelflex FMP55 5-point linearity protocol The following notes must be taken into account if option F4 ("5 point linearity protocol") has been selected in feature 550 ("Calibration"). The five points of the linearity protocol are evenly distributetd across the measuring range (0% to 100%).
  • Page 85: Customized Parametrization

    Levelflex FMP55 Customized parametrization If the option IJ "Customized parametrization HART", IK "Customized parametrization PA" or IL "Customized parametrization FF" has been selected in feature 570 "Service", customer specific presettings can be selected for the following parameters: Parameter Communication Selection list / range of values Setup →...
  • Page 86: Application Packages

    Levelflex FMP55 Application Packages Heartbeat Diagnostics Availability Available in all device versions. Function • Continuous self-monitoring of the device. • Diagnostic messages output to – the local display. – an asset management system (e.g. FieldCare/DeviceCare). – an automation system (e.g. PLC).
  • Page 87: Heartbeat Verification

    Levelflex FMP55 Heartbeat Verification Availability Available for the following options of feature 540 "Application Package": • EH: Heartbeat Verification + Monitoring • EJ:   Heartbeat Verification Device functionality checked on demand • Verification of the correct functioning of the measuring device within specifications.
  • Page 88: Heartbeat Monitoring

    Levelflex FMP55 Heartbeat Monitoring Availability Available for the following options of feature 540 "Application Package": EH: Heartbeat Verification + Monitoring Function • In addition to the verification parameters, the corresponding parameter values are also logged. • Existing measured variables, such as the echo amplitude, are used in the Foam detection and Build-up detection wizards.
  • Page 89: Accessories

    Levelflex FMP55 Accessories Device-specific accessories Weather protection cover Accessory Description Weather protection cover A0015466 300 (11.8) 275 (10.8) 255.1 (10) 164 (6.46) A0015472  40 Weather protection cover; Dimensions: mm (in)  The weather protection cover can be ordered together with the device (product structure, feature 620 "Accessory Enclosed", option PB "Weather Protection Cover").
  • Page 90 Levelflex FMP55 Mounting bracket for the electronics housing Accessory Description Mounting bracket for the electronics housing ø42...60 (1.65...2.36) 127...140 122 (4.8) (5...5.51) 161 (6.34) 162...175 (6.38...6.89) A0014793  41 Mounting bracket for the electronics housing; Dimensions: mm (in) Wall mounting Pipe mounting ...
  • Page 91 Levelflex FMP55 Remote display FHX50 Accessories Description Remote display FHX50 A0019128 • Material: – Plastic PBT – 316L/1.4404 – Aluminum • Degree of protection: IP68 / NEMA 6P and IP66 / NEMA 4x • Suitable for display modules: – SD02 (push buttons) –...
  • Page 92 Levelflex FMP55 Overvoltage protection Accessory Description Overvoltage protection for 2-wire-devices OVP10 (1 channel) OVP20 (2 channel) A0021734 Technical data • Resistance per channel: 2 * 0.5 Ω • Threshold DC voltage: 400 to 700 V • Threshold impulse voltage: < 800 V •...
  • Page 93 Levelflex FMP55 Bluetooth module for HART devices Accessory Description Bluetooth module A0036493 • Quick and easy commissioning via SmartBlue (app) • No additional tools or adapters required • Signal curve via SmartBlue (app) • Encrypted single point-to-point data transmission (tested by Fraunhofer institue) and password protected communication via Bluetooth®...
  • Page 94: Communication-Specific Accessories

    For details refer to Technical Information TI00404F Accessory Description Commubox FXA291 Connects Endress+Hauser field devices with CDI interface (= Endress+Hauser Common Data Interface) to the USB interface of a computer. Order code: 51516983  For details refer to Technical Information TI00405C...
  • Page 95: Service-Specific Accessories

    Fieldgate FXA42, FXA30 and FXA30B. SupplyCare Hosting is offered as a hosting service (Software as a Service, SaaS). In the Endress+Hauser portal, the user is provided with the data over the Internet. ...
  • Page 96: Documentation

    For an overview of the scope of the associated Technical Documentation, refer to the following: • The W@M Device Viewer : Enter the serial number from the nameplate (www.endress.com/deviceviewer) • The Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the 2-D matrix code (QR code) on the nameplate. Standard documentation...
  • Page 97: Safety Instructions (Xa)

    Levelflex FMP55 Safety Instructions (XA) Depending on the approval, the following Safety Instructions (XA) are supplied with the device. They are an integral part of the Operating Instructions. Feature 010 Approval Available for Feature 020: "Power Supply; Output" ATEX II 1G Ex ia IIC T6 Ga...
  • Page 98 Levelflex FMP55 Feature 010 Approval Available for Feature 020: "Power Supply; Output" NEPSI Ex ia IIC T6 Ga/Gb, Ex iaD 20/21 T85…90°C FMP55 XA00638F XA00638F XA00638F XA00644F XA00638F NEPSI Ex d[ia] IIC T6 Ga/Gb, DIP A20/21 T85…90°C FMP55 XA00639F XA00639F XA00639F XA00645F XA00639F IP66 FM/CSA IS+XP Cl.I,II,III Div.1 Gr.A-G...
  • Page 99 Levelflex FMP55 Ex-marking in case of connected FHX50 remote display If the device is prepared for the remote display FHX50 (product structure: feature 030: Display, Operation", option L or M), the Ex marking of some certificates changes according to the following table Feature 010 ("Approval") Feature 030 ("Display, Operation") Ex-marking...
  • Page 100 *71394761* 71394761 www.addresses.endress.com...