Run/Stop Options

Stop Field

To start the oscilloscope running in the single-mode after it has been running
in repetitive mode, touch the green Run field. Slide your finger down to the
Single field in the pop-up until the Single field turns white, then remove your
finger from the screen. This starts the oscilloscope running in the Single
mode. When single-shot acquisition is in process, the Run/Stop field will
display Stop. When the acquisition is complete, the field display changes to
Run. In Single mode, the oscilloscope makes a single acquisition and
displays the results. The oscilloscope then waits until the Run field is
touched again before making another acquisition.
Repetitive Mode Option
Repetitive mode acquisition fills acquisition memory with 32768 samples of
the input waveform on continuing acquisitions. The display is updated each
time a new acquisition is made. Repetitive mode continues acquiring data in
this manner until you touch the Stop field. As in single mode, each
32768-sample waveform record is acquired in a single acquisition.
To start the oscilloscope running in the repetitive mode after it has been
running in single mode, touch the green Run field. Slide your finger down to
the Repetitive field in the pop-up until the Repetitive field turns white, then
remove your finger from the screen. This starts the oscilloscope running in
the Repetitive mode. While the real-time repetitive acquisitions are being
made, the Run/Stop field will display Stop. When you want to stop making
repetitive acquisitions, touch the Stop field. To resume making repetitive
acquisitions, touch the Run field again.
Cancel Option
If you select the Cancel field in the pop-up, the oscilloscope returns to the
state it was in before the Run field was touched. That is, it is stopped in
either single or repetitive mode, whichever mode was selected last.
Stop Field
When the red Stop field is displayed, the oscilloscope is running. Touch the
Stop field to cause the oscilloscope to stop running.