8. XR 3500/3700 Initial Power-up and Operational Test
The system should be fully set up according to the System Setup Section
before starting these procedures.
1. To turn on the power, turn the key to the ON position. The key is
required to turn on the power and cannot be removed unless the
key has been returned to the off position. The power light should
blinks when the key switch is on and the cover lock is disengaged.
Then when the cover lock is engaged, the power light turns steady.
2. Make sure that there is nothing inside the inspection chamber,
close the cover if open and engage the lock, and depress and hold
the foot pedal. The X-ray will not come on if the cover is not locked
into position. The image on the PC or video monitor should change
from a flat light gray image to a brighter, slightly grainy image. This
is the blank x-ray image.
3. Release the foot pedal and the image should return to its original
state. To test the safety interlock circuit, disengage the cover lock,
but DO NOT OPEN THE COVER. Now press the foot pedal and
the red X-ray on light should not illuminate.
Composite Video Connection
10