Terminals (IS version)

1 – +V DC
The positive power supply from the integrator connection.
2 – Speed Out
The positive output connection of the measurement loop.
GND – Ground
A ground connection. Do not use this ground for the cable shield.
Notes:
Ground the cable shield at the integrator end only!
For optimal performance the housing of the Speed Sensor must be connected
to a reliable earthed ground.

Terminal Connections to Siemens Milltronics Integrators

WS300 IS
1
2

Terminal Connections to SIWAREX FTC Integrator

WS300 IS
1
2
Connect CI- to Common.
Page 14
SITRANS WS300 Speed Sensor – OPERATING INSTRUCTIONS 7ML19985ML01
IS Switch Isolator
Terminal
1
3
7
8
IS Switch Isolator
Terminal
1
3
7
8
Integrator
speed signal input
- excitation
FTC
CI+
IL+