Inspection Of A Unit Fitted With A Flow Housing - Honeywell Searchpoint Optima Plus Operating Instructions Manual

With optional hart output
Hide thumbs Also See for Searchpoint Optima Plus:
Table of Contents

Advertisement

MAN0905_Issue 7_07/17
SHC1 Handheld Interrogator
8. Replace the Dust Barrier and Standard Weather Protection.
9. Press Escape then Enter to release the 4-20 mA output signal.
10. Press Escape to return to the main menu.
11. Select Diagnose then Soft Reset to clear any faults or warnings that may have been generated by
cleaning.
Note: If any configuration changes have been made, wait 15 seconds before doing a soft reset or cycling
the power.
12. Press Escape to return to the main menu.

10.3 Inspection of a unit fitted with a Flow Housing

Searchpoint Optima Plus should be inspected periodically and cleaned if necessary. The required frequency
of inspection will depend on local site conditions and will need to be established through experience.
Caution: Gas readings may be generated during this procedure.
1. Select Calibrate then Display to inhibit the 4-20 mA output signal.
2. Inspect the Searchpoint Optima Plus unit, its junction box and cabling for any signs of physical damage.
3. Disconnect any pipework attached to the flow housing.
4. Using a ½ inch / 13 mm A/F open ended spanner, remove the two black nuts which retain the flow
housing. Keep the O-ring seals which are below the retaining nuts.
5. Slide the flow housing off the Searchpoint Optima Plus. This may require some force due to the friction
of the O-ring.
6. Clean any dust or contaminants from the optics using mild detergent and a soft cloth or cotton bud.
Do not use solvents or abrasive cleaners.
7. Ensure that the large sealing O-ring in the body of the flow housing, and the two small O-rings which go
below the retaining nuts, are clean and in good condition. Replace if in doubt.
8. Slide the flow housing over the end of the Searchpoint Optima Plus and push until the flow housing is
firmly seated.
9. Re-fit the o-rings and retaining nuts.
10. Press Escape then Enter to release the 4-20 mA output signal.
11. Press Escape to return to the main menu.
12. Select Diagnose then Soft Reset to clear any faults or warnings that may have been generated by
cleaning.
Note: If any configuration changes have been made, wait 15 seconds before doing a soft reset or cycling
the power.
13. Press Escape to return to the main menu.
10. Maintenance using
Searchpoint Optima Plus with optional HART
43
output
®
Operating Instructions
2108M0550