Chapter 5 Reprocessing the Endoscope (and related reprocessing accessories)
Detach accessories from the endoscope
Attach the water resistant cap (MH-553)
48
5.
Switch the airflow regulator on the light source to maximum output ("HIGH"
or "3").
6.
Depress the button of the AW channel cleaning adapter to flush the air
channel with water from the water container for 30 seconds.
7.
Release the button to flush air for 10 seconds.
8.
Turn the light source OFF.
1.
Detach the videoscope cable (MAJ-1430, MAJ-843, or MH-976) from the
electrical connector of the endoscope.
2.
Detach the suction tube from the suction connector on the endoscope
connector.
3.
Detach the metal tip of the water container (MAJ-901 or MH-884) from the
air/water supply connector on the endoscope connector. Put the metal tip of
the water container tube into the receptacle on the lid of the water container,
as described in the instruction manual for the water container.
If the exterior surface of the electrical connector of the
endoscope is scratched, the connection with the water
resistant cap (MH-553) may no longer be waterproof and the
cap's seal may be damaged. If the electrical connector is
damaged, send the endoscope to Olympus for repair.
The water resistant cap can be attached in two different
positions. Attach the water resistant cap to the endoscope as
shown in Figure 5.5 to allow the endoscope and cap to lay
properly in the reprocessing basin.
EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL