Fluke 9133 User Manual page 37

Infrared thermometer calibrator
Table of Contents

Advertisement

Controller Communications Commands continued
Command Description
Set serial linefeed mode to
on
Set serial linefeed mode to
off
Calibration Menu
Read R0 calibration
parameter
Set R0 calibration parame-
ter to n
Read ALPHA calibration
parameter
Set ALPHA calibration pa-
rameter to n
Read DELTA calibration
parameter
Set DELTA calibration
parameter
Read BETA calibration
parameter
Set BETA calibration
parameter
Functions not on menu
Read firmware version
number
Read structure of all
commands
Read all operating
parameters
Legend:
Note:
Command
Command
Format
Example
lf[eed]=on
lf=on
lf[eed]=of[f]
lf=of
r[0]
r
r[0]=n
r=100.324
al[pha]
al
al[pha]=n
al=0.003843
3
de[lta]
de
de[lta]=n
de=1.3742
be[ta]
be
be[ta]=n
be=0.342
*ver[sion]
*ver
h[elp]
h
all
all
[] Optional Command data
{} Returns either information
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
When DUPLEX is set to FULL and a command is sent to READ, the command is re-
turned followed by a carriage return and linefeed. Then the value is returned as indi-
cated in the RETURNED column.
Digital Communication Interface
Returned
Returned
Example
r0: 999.999
r0: 100.578
al: 9.9999999
al:
0.0038573
de:9.99999
de: 1.507
be:9.999
be:0.342
ver.9999,9.99
ver.9133,1.0
0
list of commands
list of parameters
8
Acceptable
Values
90 to 110
.002 to .005
0–3.0
–100.0 to
100.0
33