1.2.2
HIGH VOLTAGE is used in the operation of this equipment. SEVERE
INJURY OR DEATH may result if personnel fail to observe safety pre-
cautions. Before working inside the equipment, turn the power off and
disconnect the power cord.
Always replace the fuse with one of the same rating, voltage, and type.
Allow at least six inches of space between the instrument and nearby
objects. DO NOT place instrument under a flammable structure.
DO NOT use this unit for any application other than calibration work.
DO NOT use this unit in environments other than those listed in the
user's guide.
DO NOT operate near flammable materials.
Use of this instrument at HIGH TEMPERATURES for extended periods
of time requires caution.
Completely unattended high temperature operation is not recom-
mended for safety reasons.
Before initial use, after transport, and anytime the instrument has not
been energized for more than 10 days, the calibrator must be ener-
gized for a dry-out period of 1 to 2 hours before it can be assumed to
meet all of the safety requirements of the IEC1010-1.
The instrument can generate extreme temperatures. Precautions must
be taken to prevent personal injury or damage to objects.
Use only a grounded AC mains supply of the appropriate voltage to
power the instrument. Refer to Section 3.1, Specifications for power
details.
The instrument is equipped with operator accessible system fuses. If a
fuse blows, it may be due to a power surge or failure of a component.
Replace the fuse once. If the fuse blows a second time, it is likely
caused by failure of a component part. If this occurs, contact an Autho-
rized Fluke Hart Scientific Service Center (see Section 1.3). Always re-
place the fuse with one of the same rating, voltage, and type. Never
replace the fuse with one of a higher current rating.
This instrument is intended for indoor operation only.
Follow all safety guidelines listed in the user's manual.
Calibration Equipment should only be used by Trained Personnel.

Cautions

Use the target cover at temperatures below 5°C. If ice forms on the tar-
get, the IR probe will not indicate the correct temperature.
DO NOT use fluids to clean out the target surface.
1