4.4
MAG 5000/6000 CT
To ensure that the settings of this custody transfer-approved MAG 5000/6000 CT transmitter
are not changed, it is necessary to install a hardware key to lock the software functions and
to seal the device.
MAG 6000 CT is installed like a Standard MAG 6000 except for the final sealing.
Calibration sealing has been carried out at calibration.
4.4.1
Installing hardware key
Use hardware key on non-verified transmitter
1. Mount hardware key on transmitter connection plate during setting of primary operating
2. Remove hardware key after setting up and calibrating unit.
SITRANS F M MAG 5000/6000
Operating Instructions, 01/2010, SFIDK.PS.026.G1.02
parameters such as Q
max
commissioning or calibration. See setup menus in appendix menu diagrams (Page 73).
This locks the menu structure and the selected settings.
Note
Hardware key function
Setting of primary operating parameters is blocked during normal operation.
When key is mounted, access to all menu items is gained. When key is removed, primary
settings are blocked in accordance with requirements in authorisation.
., low-flow cut-off, units, approvals, etc. in connection with
Installing/Mounting
4.4 MAG 5000/6000 CT
27