Guided Wave Radar
SITRANS LG270
4 ... 20 mA/HART - four-wire
Rod and cable probe
-196 ... +280 °C; -196 ... +450 °C
Operating Instructions • 09/2017

Advertisement

Table of Contents
loading

Summary of Contents for Siemens SITRANS LG270

  • Page 1 Guided Wave Radar SITRANS LG270 4 … 20 mA/HART - four-wire Rod and cable probe -196 … +280 °C; -196 … +450 °C Operating Instructions • 09/2017...
  • Page 2 SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 3: Table Of Contents

    Set up with other systems ....................56 DD adjustment programs ....................56 Field Communicator 375, 475 ..................56 Diagnostics and servicing ....................57 Maintenance ........................57 Diagnosis memory ......................57 Status messages ......................58 Rectify faults ........................61 PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 4 11.2 Dimensions ........................81 11.3 Trademark ........................85 Safety instructions for Ex areas Take note of the Ex specific safety instructions for Ex applications. These instructions are attached as documents to each instrument with Ex approval and are part of the operating instructions manual. Editing status: 2017-09-14 SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 5: About This Document

    This arrow indicates a single action. Sequence of actions Numbers set in front indicate successive steps in a procedure. Battery disposal This symbol indicates special information about the disposal of bat- teries and accumulators. PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 6: For Your Safety

    During work on and with the device the required personal protective equipment must always be worn. Appropriate use SITRANS LG270 is a sensor for continuous level measurement. You can find detailed information about the area of application in chapter "Product description". Operational reliability is ensured only if the instrument is properly used according to the specifications in the operating instructions manual as well as possible supplementary instructions.
  • Page 7: Eu Conformity

    English language. Installations in the US shall comply with the relevant requirements of the National Electrical Code (ANSI/NFPA 70). Installations in Canada shall comply with the relevant requirements of the Canadian Electrical Code PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 8: Product Description

    Standard electronics with operating voltage 90 … 253 V AC; 50/60 Hz: Typ FX80B.- • Standard electronics with supply voltage 9.6 … 48 V DC; 20 … 42 V AC: Type FX80I.- Scope of delivery The scope of delivery encompasses: • Sensor • Optional accessory • Documentation SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 9: Principle Of Operation

    Principle of operation Application area The SITRANS LG270 is a level sensor with cable or rod probe for continuous level or interface measurement, particularly suitable for applications in high temperatures up to +450 °C (842 °F). Functional principle -...
  • Page 10 Example: upper medium dielectric constant 2, lower medium at least dielectric constant 12. Gas phase (L3) • Air or gas mixture • Gas phase - dependent on the application, gas phase does not always exist (d2 = 0) SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 11: Packaging, Transport And Storage

    "LG Local Display Interface" (Document-ID 43838). External display and The LG Remote Interface is an external display and adjustment unit adjustment unit for sensors with single chamber housing and Ex-d double chamber housing. PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 12 Curved segments: 100 x 100 mm (3.94 … 3.94 in) You can find further information in the operating instructions manual "Rod and cable components SITRANS series LG". Centering If you mount the SITRANS LG270 in a bypass tube or standpipe, you have to avoid contact to the bypass tube by using a spacer at the probe end. You can find additional information in the operating instructions manual "Centering".
  • Page 13: Mounting

    Make sure before mounting that all parts of the instrument exposed to conditions the process are suitable for the existing process conditions. These are mainly: • Active measuring component • Process fitting • Process seal Process conditions in particular are: PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 14: Mounting Instructions

    Mounting instructions Installation position Mount SITRANS LG270 in such a way that the distance to vessel installations or to the vessel wall is at least 300 mm (12 in). In non- metallic vessels, the distance to the vessel wall should be at least 500 mm (19.7 in).
  • Page 15 In such cases, always carry out a false signal suppression after mounting. You can find further information under "Setup procedure". ≤ 150 mm ( 5.91") DN40 ... DN150 ≤ 100 mm (3 .94") > DN150 ... DN200 Fig. 6: Mounting socket When welding the socket, make sure that the socket is flush with the vessel top. PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 16 By doing this, you avoid damage to the electronics through inductive coupling. Inflowing medium Do not mount the instruments in or above the filling stream. Make sure that you detect the product surface, not the inflowing product. Fig. 8: Mounting of the sensor with inflowing medium Measuring range The reference plane for the measuring range of the sensors is the sealing surface of the thread or flange. SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 17 If durability is no problem, we recommend the use of uncoated metal standpipes. When the SITRANS LG270 is used in bypass tubes, contact with the tube wall must be avoided. We recommend for this purpose a cable probe with centering weight.
  • Page 18 For standpipes, select the probe length such that the upper blocking distance (dead band) of the probe is above the upper ventilation hole. This allows you to measure the total level range of the medium in the SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 19 If durability is no problem, we recommend the use of uncoated metal standpipes. When the SITRANS LG270 is used in standpipes, contact with the tube wall must be avoided. We recommend for this purpose a cable probe with centering weight.
  • Page 20 Instructions for the measurement: • The 100 % point with standpipes should be below the upper ventilation hole. • The 0 % point in standpipes should be above the gravity or center- ing weight. SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 21 Make sure that the probe cable is not completely taut. Avoid tensile loads on the cable. Avoid undefined vessel connections, i.e. the connection must be either grounded reliably or isolated reliably. Any undefined change of this condition can lead to measurement errors. If there is a danger of the rod probe touching the vessel wall, fasten the probe at the bottom end. PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 22 Keep in mind that measurement is only possible up to the tensioning component. For this reason, order the cable probe 270 mm longer. L = L1 + 270 mm (10.63 in) SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 23 Rod extension In case of difficult installation conditions, for example in a socket, the probe can be suitably adapted with a rod extension. To compensate for the resulting changes in signal runtime, let the instrument determine the probe length automatically. You can find further information in the supplementary instructions of the rod and cable components. PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 24: Connecting To Power Supply

    In the case of instrument housings with metric thread, the cable glands are screwed in at the factory. They are sealed with plastic plugs as transport protection. You have to remove these plugs before electrical connection. SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 25: Connecting

    When reinserting the terminal block, you should hear it snap in. Connection procedure Proceed as follows: 1. Unscrew the housing lid 2. Loosen compression nut of the cable gland and remove blind plug PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 26 The terminal blocks are pluggable and can be removed from the housing insert. To do this, lift the terminal block with a small screwdriv- er and pull it out. When inserting the terminal block again, you should hear it snap in. SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 27: Wiring Plan, Double Chamber Housing

    Terminal Function Polarity Voltage supply Voltage supply 4 … 20 mA output (active) 4 … 20 mA output (passive) Mass - output Function ground when in- stalling according to CSA (Canadian Standards Asso- ciation) PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 28: Supplementary Electronics

    "Additional current output" First current output (I) - Voltage supply and signal output, sensor (HART) Additional current output (II) - Voltage supply and signal output (without HART) Ground terminal for connection of the cable screen SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 29: Switch-On Phase

    As soon as a plausible measured value is found, the corresponding current is outputted to the signal cable. The value corresponds to the actual level as well as the settings already carried out, e.g. factory setting. PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 30: Set Up With The Display And Adjustment Module

    Fig. 19: Insertion of the display and adjustment module with single chamber housing Note: If you intend to retrofit the instrument with a display and adjustment module for continuous measured value indication, a higher lid with an inspection glass is required. SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 31: Adjustment System

    5 s, the display returns to the main menu. The menu language is then switched over to "English". Approx. 60 minutes after the last pressing of a key, an automatic reset to measured value indication is triggered. Any values not confirmed with [OK] will not be saved. PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 32: Parameter Adjustment - Quick Setup

    Switch-on phase After switching on, the SITRANS LG270 carries out a short self-test where the device software is checked. The output signal transmits a fault signal during the switch-on phase. The following information is displayed on the display and adjustment module during the startup procedure: •...
  • Page 33 "->" key you jump to the next position. You can enter names with max. 19 characters. The character set comprises: • Capital letters from A … Z • Numbers from 0 … 9 • Special characters + - / _ blanks PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 34 In this menu item, you can define the type of medium (product). Setup - Application - Me- dium, dielectric constant This menu item is only available if you have selected level measure- ment under the menu item "Application". SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 35 In this menu item you can enter the max. adjustment for the level. With Level interface measurement this is the maximum total level. Adjust the requested percentage value with [+] and store with [OK]. PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 36 As an alternative, you have the possibility taking over the adjustment of the level measurement also for the interface. Enter the respective distance value in m for the surface of the upper medium corresponding to the percentage value. SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 37 In the following, you have to enter the values for your vessel, for example the vessel height and the socket correction. PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 38 In the menu item "Current output mode" you determine the output mode characteristics and reaction of the current output in case of fault. The default setting is output characteristics 4 … 20 mA, fault mode < 3.6 mA. SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 39 The level would then no longer be detectable in this area. If a false signal suppression has already been created in the sensor, the following menu window appears when selecting "False signal suppression": PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 40 The following submenu points are available: The submenu points are described below. Display - Menu language This menu item enables the setting of the requested national lan- guage. SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 41 In delivery status, the lighting is switched on. Diagnostics - Device In this menu item, the device status is displayed. status When the instrument displays a failure message, you can here get detailed information on the failure reason. PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 42 The respective min. and max. measured values are saved in the ues, Additional sensor. The values are displayed in the menu item "Peak values Ad- ditional". This menu item displays the peak values of the electronics tempera- ture as well as the dielectric constant. SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 43 Select the requested simulation variable and set the requested value. Caution: During simulation, the simulated value is outputted as 4 … 20 mA cur- rent value and digital HART signal. PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 44 In this menu item, the internal clock of the sensor is set. Time After a reset, certain parameter adjustments made by the user are Additional settings - Reset reset. Note: After this menu window, the reset process is carried out. No further safety inquiry follows. SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 45 Distance: Probe length - take dead band into account Max. adjustment - Interface 100 % Distance: 0.000 m(d) - note block- ing distances Min. adjustment - Interface Distance: Probe length - take dead band into account PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 46 20.5 mA Menu - Display Menu Menu item Default value Display Language Selected language Displayed value 1 Filling height Displayed value 2 Electronics temperature Display format 1 Automatically Display format 2 Automatically Backlight Switched on SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 47 We recommend to save the instrument adjustments. In case of an electronics exchange the saved parameter adjustment data relieve this process. Additional settings - Scal- Since scaling is very extensive, scaling of the level value was divided ing level into two menu items. PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 48 0 % and format 100 %. Additional settings - Cur- Since scaling is very extensive, scaling of the level value was divided rent output into two menu items. SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 49 The default setting is "Analogue current output" and the address 00. Additional settings - Spe- In this menu item you gain access to the protected area where cial parameters you can enter special parameters. In exceptional cases, individual parameters can be modified in order to adapt the sensor to special requirements. PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 50: Saving The Parameterisation Data

    The following data or settings for adjustment of the display and ad- justment module are saved: • All data of the menu "Setup" and "Display" • The items "Sensor-specific units, temperature unit and linearisa- tion" in the menu "Additional settings". SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 51 If it is necessary to exchange a sensor, the display and adjustment module is inserted into the replacement instrument and the data are likewise written into the sensor via the menu item "Copy device settings". PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 52: Setup With Pactware

    Connection via HART Fig. 23: Connecting the PC via HART to the signal cable SITRANS LG270 2 HART resistance approx. 250 Ω (optional depending on processing) Connection cable with 2 mm pins and terminals Processing system/PLC HART modem Voltage supply Necessary components: SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 53: Parameter Adjustment With Pactware

    You can save and print your project documentation as well as import and export projects. You can also save measured value and echo curves in the DTM. Furthermore a tank calculation program as well as a multiviewer for PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 54: Set Up With The Quick Setup

    Extended adjustment Maintenance Quick setup With quick setup you can carry out the parameter adjustment of SITRANS LG270 for your application in just a few simple steps. The assistant-driven adjustment includes the basic settings for simple, reliable setup and commissioning. Information: If the function is inactive, then possibly no instrument is connected.
  • Page 55: Saving The Parameterisation Data

    Click to the button "Quick setup", to start the assistant-driven adjust- ment for a simplified and reliable setup. Saving the parameterisation data We recommend documenting or saving the parameterisation data via PACTware. That way the data are available for multiple use or service purposes. PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 56: Set Up With Other Systems

    This software is updated via the Internet and new EDDs are automatically accepted into the device catalogue of this software after they are released by the manufacturer. They can then be transferred to a Field Communicator. SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 57: Diagnostics And Servicing

    Echo curve of the setup: This is used as reference echo curve for the measurement conditions during setup. Changes in the measure- ment conditions during operation or buildup on the sensor can thus be recognized. The echo curve of the setup is stored via: • PC with PACTware/DTM • Control system with EDD PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 58: Status Messages

    This status message is inactive by default. It can be activated by the user via PACTware/DTM or EDD. Maintenance: Due to external influences, the instrument function is limited. The measurement is affected, but the measured value is still valid. Plan in maintenance for the instrument because a failure is expected in the near future (e.g. due to buildup). SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 59 Error during setup Carry out a reset Bit 9 of Byte 0 … 5 • • Error when carrying out a Repeat setup Error in the reset instrument • False signal suppression settings faulty PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 60 • Use coaxial probe • • S602 Compensation echo super- 100 % adjustment: Increase Bit 10 of Byte 14 … 24 imposed by medium value Level with- in the search range, com- pensation echo SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 61: Rectify Faults

    The operator of the system is responsible for taking suitable meas- tion occurs ures to rectify faults. Procedure for fault recti- The first measures are: fication • Evaluation of fault messages via the adjustment device • Checking the output signal • Treatment of measurement errors PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 62 Wherever the sensor displays a constant value, the reason could also be the fault setting of the current output to "Hold value" • If the level indication is too low, the reason could be a line resist- ance that is too high SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 63 • 0 m distance signals in the close range. The Check installation conditions • sensor goes into overfill protec- If possible, switch off the func- tion mode. The max. level (0 m tion "Overfill protection" time distance) as well as the status message "Overfill protection" are outputted. PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 64: Exchanging The Electronics Module

    Internet (see operating instructions manual "Electronics module"). Caution: All application-specific settings must be entered again. That's why you have to carry out a fresh setup after exchanging the electronics. If you saved the parameter settings during the first setup of the sen- sor, you can transfer them to the replacement electronics module. A fresh setup is then not necessary. SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 65: Exchange Or Shorten Cable/Rod

    4. Screw the new rod and the new cable manually to the thread on the process fitting. 5. Exert counterforce with the second fork spanner and tighten the measuring rod or cable on the flat surfaces with a torque of 20 Nm (15 lbf ft). Fig. 36: Exchange cable or rod Information: Please maintain the specified torque so that the max. tensile strength of the connection remains. PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 66: Software Update

    C Centering weight - cable ø 2 mm Threaded pins Thread M8 for eye-bolt Fixing screw - centering weight Software update The following components are required to update the sensor soft- ware: • Sensor • Voltage supply SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 67: How To Proceed If A Repair Is Necessary

    Instruments with approvals can be bound to certain software versions. Therefore make sure that the approval is still effective after a software update is carried out. You can find detailed information in the download area on our home- page: www.siemens.com/sitranslg. How to proceed if a repair is necessary If it is necessary to repair the instrument, please contact Sie- mens Milltronics Process Instruments. You find the locations on "www.siemens.com/sitranslg". PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 68: Dismount

    Pass the instrument directly on to a spe- cialised recycling company and do not use the municipal collecting points. These may be used only for privately used products according to the WEEE directive. SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 69: Supplement

    Ʋ The Second Line of Defense (SLOD) is a second level of the process separation in the form of a gas-tight feedthrough in the lower part of the housing, preventing product from penetrating into the housing. Ʋ Supporting material 316L PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 70 450 Nm (332 lbf ft) Ʋ -196 … +450 °C (-321 … +842 °F) max. 400 Nm (295 lbf ft) Torque for exchangeable cable or rod probe (in the process fitting) Ʋ Cable: ø 2 mm (0.079 in) 20 Nm (14.75 lbf ft) SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 71 Ʋ Relative humidity 45 … 75 % Ʋ Air pressure +860 … +1060 mbar/+86 … +106 kPa (+12.5 … +15.4 psig) The output values can be assigned individually. The indication values can be assigned individually. PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 72 Probe end does not touch the vessel bottom Sensor parameter adjustment No gating out of false signals carried out Fig. 38: Measuring ranges - SITRANS LG270 Reference plane Probe length L Measuring range (default setting refers to the measuring range in water)
  • Page 73 -5 mm (-0.197") 0,08 m 0,3 m 0,015 m (3.15") (11.811") (0.591") Fig. 39: Deviation SITRANS LG270 in rod version in water Dead band (no measurement possible in this area) Probe length 25 mm (0.984") 2 mm (0.079") -2 mm (0.079")
  • Page 74 (5.906 (7.874 " ") (3.937 0,08 m ") (3.15 ") Fig. 42: Deviation SITRANS LG270 in cable version (2 mm/0.079 in), in medium oil Dead band (no measurement possible in this area) Probe length SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 75 0,2 m 0,1 m (5.906") (7.874") (3.937") 0,15 m (5.906") Fig. 43: Deviation SITRANS LG270 in cable version (4 mm/0.157 in), in medium oil Dead band (no measurement possible in this area) Probe length Repeatability ≤ ±1 mm Variables influencing measurement accuracy Specifications for the digital measured value Temperature drift - Digital output ±3 mm/10 K relating to the max.
  • Page 76 -1 … +400 bar/-100 … +40000 kPa (-14.5 … +5800 psig), depending on the process fitting Process temperature -196 … +280 °C (-321 … +536 °F) Time span after a sudden measuring distance change by max. 0.5 m in liquid applications, max 2 m with bulk solids applications, until the output signal has taken for the first time 90 % of the final value (IEC 61298-2). SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 77 -40°C / -40°F Fig. 45: Ambient temperature - process temperature, standard version Ambient temperature Process temperature (depending on the seal material) Aluminium housing Plastic housing Stainless steel housing, precision casting Stainless steel housing, electropolished PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 78 Flanges of 904L (1.4539): see ASME B16.5-2013, Table 2-3.11, permissible temperature range: -60 … +400 °C (-76 … 752 °F) Vibration resistance Ʋ Rod probe 1 g with 5 … 200 Hz according EN 60068-2-6 (vibration at resonance) with rod length 50 cm (19.69 in) SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 79 Ʋ Analogue Via the current output Ʋ Digital Via the digital output signal (depending on the electron- ics version) Range -40 … +85 °C (-40 … +185 °F) Resolution < 0.1 K Accuracy ±3 K PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 80 Instruments with approvals can have deviating technical data (depending on the version). For such instruments, the corresponding approval documents must be noted. The prerequisites for maintaining the protection rating are a suitable cable as well as correct mounting. When used with fulfilled housing protection. SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 81: Dimensions

    11.2 Dimensions Housing ~ 81 mm (3.19") ø 77 mm (3.03") M16x1,5 M20x1,5/ ½ NPT Fig. 48: Dimensions of housing - with integrated display and adjustment module the housing is 9 mm/0.35 inches higher PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 82 (0.79") (0.79") Fig. 49: SITRANS LG270, threaded version with gravity weight (all gravity weights with thread M8 for eye-bolt) Sensor length, see chapter "Technical data" Cable ø 4 mm (0.157 in), temperature version -196 … +280 °C (-321 … 536 °F) Cable ø...
  • Page 83 ø 4 mm (0.16") (0.08") (0.08") (0.16") Fig. 50: SITRANS LG270, threaded version with centering weight Sensor length, see chapter "Technical data" ø 40 mm (1.57 in) ø 45 mm (1.77 in) ø 75 mm (2.95 in) ø 95 mm (3.74 in) Cable ø...
  • Page 84 ø 16mm (0.63") (0.63") Fig. 51: SITRANS LG270, threaded version Sensor length, see chapter "Technical data" Rod ø 16 mm (0.63 in), temperature version -196 … +280 °C (-321 … 536 °F) Rod ø 16 mm (0.63 in), temperature version -196 … +450 °C (-321 … 842 °F) Max.
  • Page 85: Trademark

    11.3 Trademark All the brands as well as trade and company names used are property of their lawful proprietor/ originator. PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 86 Echo curve of the setup 44 EDD (Enhanced Device Description) 56 Quick setup 32 Electrical connection 25 Electronics compartment - double chamber housing 27 Read out info 50 Error codes 60 Repair 67 Event memory 57 Replacement parts SITRANS LG270 - Operating Instructions PBD-51041064...
  • Page 87 Reset 44 Scaling measured value 47, 48 Sensor characteristics 50 Sensor status 41 Simulation 43 Special parameters 49 Type label 8 Type of medium 34 Units 34 Vessel insulation 21 Voltage supply 24, 80 PBD-51041064 SITRANS LG270 - Operating Instructions...
  • Page 88 For more information www.siemens.com/level www.siemens.com/weighing Siemens Canada Limited Subject to change without prior notice PD PA PI LW PBD-51041064 Rev. 6.0 1954 Technology Drive P.O. Box 4225 © Siemens AG 2017 Peterborough, ON K9J 7B1, Canada email: techpubs.smpi@siemens.com Printed in Canada www.siemens.com/processautomation...