Integrator Calibration

Refer to the flowmeter integrator instruction manual for integrator calibration
instructions.
To apply the test weight for the span calibration:
1.
Mount the calibration bracket (stored on the left side of the cast frame).
2.
Attach one end of a string (monofilament fishing line or fine flexible cable) to the test
weight.
3.
Route the other end of string over the calibration pulley.
4.
Attach the free end of the string to the string fixing bolt. Ensure the string rests in the
bolt groove.
Note:
Ensure the test weight is suspended free from obstruction.
5.
When the calibration is complete, return the calibration bracket to the storage
position.
Note:
Accurate calibration is not assured until material tests and a span adjust
are performed.
Page 10
Milltronics ILE-61 – INSTRUCTION MANUAL
string fixing bolt
calibration arm
test weight
7ML19985CX01