7.2.3 Set 4-20 mA output "mA"
1. Use 's' up or 't' down buttons to select the set
4-20mA 'mA' submenu and press '3' to accept.
2. '4 mA' is displayed
3. Use 's' up or 't' down buttons to change the
gas conc. corresponding to 4 mA.
4. Press '3' to display 20 mA
5. Use 's' up or 't' down buttons to change the
gas conc. corresponding to 20 mA.
6. Use '3' to update all the changes (Updt
displayed) and return to step 1.
7. Press 's' up or 't' down to select another
submenu.
8. Press 'X' to return to set up, calibration and test
menu selection
9. Press 'X' again to return to normal operation.
Midas
7.2.4 Set Faults ' FLt'
1. Use the 's' up or 't' down buttons to select
the set faults ' FLt' submenu and press '3' to
accept.
2. The flashing fault relay configuration (1FLt, 2 FLt,
CmbF or nEtr) is displayed.
3. U s e t h e ' s ' u p o r ' t ' d o w n b u t t o n s
t o
c h a n g e
(See Table 7-3 for details of fault relay
configurations)
4. Press '3' to display the flashing fault output
latching (L) or non latching (nL) setting is
displayed.
5. Use the 's' up or 't' down buttons to change
between the settings.
6. Press '3' to display the flashing fault output
normally energized (nE) or normally de-energized
(nd) setting.
7. Use the 's' up or 't' down buttons to change
between the settings.
8. Press '3' to set the fault m12 frequency.
9. Use the 's' up and 't' down buttons to change
between settings:
LCD Display
Wm 12
cartridge expiration weekly reminder
0m 12
no reminder
1m12
one-time reminder
10. Press '3' to display the Temperature fault m15
and F80 enable (Y) or disable (n) setting.
11. Use the 's' up or 't' down buttons to change
between settings.
12. Press '3' to update all the changes (UPdt
displayed) and return to step 1.
13. Press 's' up or 't' down to select another
submenu.
7-9
Gas Detector
®
t h e
c o n f i g u r a t i o n .
m12 Frequency

Hide quick links: