S112=
Verbose Replies (Write Only)
Enables (Non-Zero value) or disables (0) Verbose Replies.
ENABLE VERBOSE
DISABLE VERBOSE
ENABLE RS485
DISABLE RS485
IS485TERMINATED
NOTERMINATERS485
TERMINATERS485
RESTOREFACTORYDEFAULTS
Reset values include: Product config (S64), Piezo calibration curve settings, Temperature
settings, Analog output settings
TDAO
DAC zero output
In calibration mode, generate output voltage equivalent to the DAC zero code (S51).
TDAS
DAC span output
In calibration mode, generate output voltage equivalent to the DAC full scale code (S52).
TDAZ
DAC zero count output
In calibration mode, generate output voltage equivalent to 0 counts to the DAC.
TOFF
DAC adjust complete
In calibration mode, write the D/A converter values to memory.
ZERO
Set Pressure Zero
Sets S15 UB Zero and S16 DB Zero to the current bridge power readings. The effect of this is
the current pressure reading becomes the zero pressure point.
ZRO
Alias for ZERO
For backward compatibility.
HVG-2020A Instruction Manual
Display descriptive text and units. Same as S112=1.
Display values only. Same as S112=0.
Configure the serial port for RS485 communication
Configure the serial port for RS232 communication
Read whether termination resistors are engaged
Disable termination resistors on RS485 communications
Enable termination resistors on RS485 communications
Reset variables to the values stored at the factory.
Page 22 of 29