Fitting An Adaptor Board To A Magmaster - ABB CalMaster2 Instruction Manual

Tests the integrity of a flow metering system by altering a number of flow measurement parameters and measuring the values returned from the flow measuring system.
Hide thumbs Also See for CalMaster2:
Table of Contents

Advertisement

Field and Software Tools
CalMaster2 and CheckMaster

3.2.1 Fitting an Adaptor Board to a MagMaster

Note. The newer series (VKH) MagMaster does not need to have this adaptor board fitted as
one is already incorporated into its design.
To fit an adaptor:
1.
Turn off the power to the MagMaster and remove the covers – see Fig. 3.1.
2.
Remove the wiring from the terminals. Slacken the terminal screws by at least 6 turns.
3.
Carefully ease the sensor wiring to one side and fit the adaptor board so that the extended
pins fit into the terminals – see Fig. 3.2. Tighten the terminal screws.
4.
Fit and tighten the securing screw to the corner of the adaptor board.
5.
Connect the sensor cable leads to the adaptor plug – see Fig. 3.3. Do not fit the plug to the
socket on the adaptor board.
Fig. 3.2 Fitting the Adaptor Board
Pin 1
Fig. 3.3 Fitting the Sensor Cable Plug
10
Adaptor
Board
Pin 8
Grey
DS 2
Pin 7
Pink
SIG 2
Pin 6
Blue
SIG 1
Pin 5
White
DS 1
Pin 4
Purple
SIG GRN
Pin 3
Bare
Drain (optional)
Pin 2
Yellow
CD 2
Pin 1
Red
CD 1
IM/VCM2 Issue 6