(g) – Indicate "Time to Iso" for each Sample channel. Time to Iso is the time it takes for the
calorimeter to reach isothermal conditions. "Time to Iso" varies with sample type and size and can
be estimated by running duplicate samples. When plotting Energy and Power, Time to Iso is used
as a point of origin, i.e. all values before Time to Iso are ignored in Energy and Power calculations.
See ASTM C1679 for more details.
(h) – Use Save Batch to save the current configuration of data logs, so they can be called up as on
batch using "Load Batch" in the future.
(j) – Data in area (j) can be edited. If data is edited, a yellow light will light up for the corresponding
Sample number in the Conf Changed field on the right, as shown in the picture below. The light will
switch off after you saved your changes by pressing "Save Updated Log" or "Save All Updated
Logs". Please note that changes cannot be undone once saved.
(k) - Switch between Voltages, Power or Energy data by clicking on the box on top of the graph, as
shown in the example below (Energy plot).
(i)