5890 TCD Troubleshooting Guidelines
A15956
This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use of this material.
The information contained herein is intended for use by informed individuals who can and must determine its fitness for their purpose.
a15956cover.doc
Page 1 1 1 1 of 1 1 1 1
http:// www.chem.agilent.com
http:// www.chem.agilent.com
http:// www.chem.agilent.com
http:// www.chem.agilent.com

Advertisement

loading

Summary of Contents for Agilent Technologies 5890

  • Page 1 A15956 This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use of this material. The information contained herein is intended for use by informed individuals who can and must determine its fitness for their purpose.
  • Page 2 5890 TCD Troubleshooting When trouble shooting the TCD, as in any detector. You should begin with half-splitting the detector from the column and the inlet. This will ensure that you are looking in the correct place for the problem. The first things to look/listen for: Is there an audible ticking sound coming from the switching valve? 2.
  • Page 3 G. Connections to the data recorder H. Defective electronics board NOTE. There could be other causes to any of the above symptoms. If you verify all of these first the next thing That you should do is contact Agilent Technologies Service.(1-800-227-9770)