X-Ray Generator Technical Specifications
7–6
TECHNICAL SPECIFICATIONS
Figure 7–1
Heating and Cooling Curves of the X-ray Generator.
110, 0
100, 0
90, 0
80, 0
70, 0
60, 0
50, 0
40, 0
30, 0
20, 0
10, 0
-
0
60
120
Table 7–4
Beam Limitations of the X-ray Generator
Manufacturer
Type
Maximum symmetrical field of radiation at a distance of
613 mm from the focal point (Panoramic mode)
Maximum symmetrical field of radiation at a distance of
690 mm from the focal point (3D mode)
Table 7–5
Characteristics of the X-ray Tube
Manufacturer's name
Type
Nominal high voltage
Nominal anode power
Maximum amount of heat accumulated in the anode
Nominal focal spot diameter (IEC 60336)
Anode materials
Target angle
Inherent filtration
Heating and cooling curves
180
240
300
360
420
Time in minutes
Trophy
Rigidly mounted unit with fixed window dimensions,
not removable, and integrated x-ray generator
4 (-0.5/+1) x 128 (± 2) mm
47 (± 1) x 59 (± 1) mm
CEI
OPX 105
90 kV
810 W
30 kJ
0.5 mm (0.020'')
Tungsten
0.5 mm (0.20 ") eq. Al
480
540
600
660
720
100 W
66 W
33 W
15 W
10 W
Cooling