Unit Electronic Specifications
7–4
TECHNICAL SPECIFICATIONS
Unit Electronic Specifications
Type of Electrical Power Supply
Acceptable fluctuation
Apparent resistance of the power
supply circuit
Permanent absorbed current
Current absorbed during the x-ray
emission
Maximum absorbed power
Protection for the power supply
system
Nominal high voltage
Maximum corresponding tube
current
Nominal tube current
Maximum corresponding high
voltage
Tube current/voltage combination
for maximum output power
Nominal power for an exposure time
of 0.1 s.
Utilization Rate in Continuous Mode (for
example: one exposure - 85 kV, 5 mA - 13.9
second, every 3 minutes)
33 W
Selection of the Load Parameters:
kV (in increments of 1 kV)
mA (in increments of 25%)
Cooling Conditions
Maximum dissipation of heat from the x-ray
radiogenic assembly into the ambient air (for
utilization rate in continuous mode)
Accuracy of the Load Parameters
High voltage
Current in the tube
Exposure time seconds
230/240 V (± 10%) 50/60 Hz,
Single-Phase
± 10%
Ω
0.5
(max.)
0.45 A
9.5 A
2.2 kVA
By shutter release at a maximum
current of 16A and a differential
current of 30 mA
90 kV
10 mA
15 mA
80 kV
80 kV, 15 mA
at 80 kV, 15 mA: 1200 W
Utilization Rate in Intermittent Mode (for example: one
exposure - 80 kV, 15 mA - 13.9 second, every 3 minutes)
From 60 to 90 kV
From 2 to 15 mA
33 W
kV ± 10%
mA ± 20%
Seconds ± (10% + 1ms)
100/110/130V (± 10%) 50/60 Hz,
Single-Phase
± 10%
Ω
0.12
(max.)
0.9 A
22 A
2.2 kVA
By shutter release at a maximum
current of 20A and a differential
current of 30 mA
90 kV
10 mA
15 mA
68 kV
85 kV, 12 mA
at 85 kV, 12 mA: 1020 W
93 W