Guidance and manufacturer's declaration – electromagnetic immunity (continued)
The AutoSet CS2 is intended for use in the electromagnetic environment specified below. The customer or the
user of the AutoSet CS2 should assure that it is used in such an environment.
IEC60601-1-2
Immunity test
test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
10 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
NOTE 1: At 80 MHz and 800MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the AutoSet CS2 is used
exceeds the applicable RF compliance level above, the AutoSet CS2 should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or
relocating the AutoSet CS2.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Compliance
Electromagnetic environment – guidance
level
Portable and mobile RF communications
equipment should be used no closer to any part
of the AutoSet CS2, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
d = 1.17 √P
3 Vrms
d = 0.35 √P 80 MHz to 800 MHz
10 V/m
d = 0.70 √P 800 MHz to 2.5 GHz
where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range.