NSUFFICIENT PRESSURE SUPPORT ALARM
Is triggered if for 3 breaths in a row the maximum mask pressure achieved at
inspiratory to expiratory transition is either <87.5% of the absolute mask pressure
targetted at that point, or < absolute mask pressure targetted minus 1 cmH
whichever is the lower.
H
IGH LEAK ALARM
Leak exceeds 30 L/min for more than 20 seconds.
F
LOW BLOCKED ALARM
Flow is less than 6 L/min for more than 20 seconds.
IEC 601-1
Class II, Type CF
Class II medical electrical equipment provides protection against electric shock either
by double insulation or reinforced insulation and does not require a protective earth.
Table 3: Displayed values
Value
Proximal cuff
Pressure
Mass flow sensor
Leak
Respiratory rate
Tidal volume
Minute ventilation
Target MV
Table 4: Maximum low impedance flow at stated pressures
Pressure
(cmH
4.2
8.6
13.0
44
CLASSIFICATIONS
Range
-3 to 40 cmH
0–99 litres per
minute
8–30 breaths per
minute
0–4500 millilitres
0–40 litres per
minute
0–40 litres per
minute
Flow (L/min)
O)
2
217
218
219
Accuracy
O
0.5 cmH
O
2
2
+/- 12 litres per
minute
+/- 10%
+/- 10%
+/- 10%
+/- 10%
O,
2
Display
Resolution
0.1 cmH
O
2
1 litre per breath
1 breath per
minute
1 millilitre
0.1 litres per
minute
0.1 litres per
minute