Advertisement

5
Configuration
68
Standby flow: Enter the flow rate of purge gas that flows
through the sample loop, sample probe, and vent between
sequences. This purge flow keeps the sampling system dry
and prevents atmospheric contamination from entering the
HS.
• Enter the desired flow rate, in mL/min, then press [Enter].
• Select Off to turn off the standby flow (not
recommended).
Status: Select and press [Enter] to change the order of status
items displayed.
Scroll to the first item to display in the status. Press
1
[Enter].
2
Continue selecting items.
The items will appear in the status display in the order
that you selected them.
Clock: Select to set the date and time, or to set the preferred
date format.
Scroll to Time zone and enter the offset from GMT
1
(Greenwich Mean Time), in hours and minutes (hhmm).
Press [Enter].
Scroll to Time and enter the current local time, using
2
24- hour format.
3
Scroll to format. Press [Mode/Type] to select the desired
format for displaying the date.
Scroll to Date and enter the date in the current format.
4
APG Polarity: See
"APG Remote
Instrument Schedule: See
Conservation)".
Polarity".
"Instrument Schedule (Resource