Example Of Use - Hach ORBISPHERE 410 User Manual

Hide thumbs Also See for ORBISPHERE 410:
Table of Contents

Advertisement

Configuration report example
The "Configuration" below is given for a one channel instrument.
INSTRUMENT CONFIGURATION
Measurement mode . . . . . . . . . Continuous mode
Pressure unit . . . . . . . . . . . . . . . [bar]
Temperature unit . . . . . . . . . . . . [°C]
Storage mode . . . . . . . . . . . . . . Rolling buffer
Storage RAM time . . . . . . . . . . . 10 [s]
Storage FLASH time . . . . . . . . . 3600 [s]
Channel 1
Membrane . . . . . . . . . . . . . . . . . 2956A
Medium . . . . . . . . . . . . . . . . . . . Liquid
Gas unit. . . . . . . . . . . . . . . . . . . mbar
Liquid . . . . . . . . . . . . . . . . . . . . Water
Resolution displayed . . . . . . . . . 3
Thermal cut off . . . . . . . . . . . . . Enabled60.0 [°C]
Alarms
Low Low . . . . . . . . . . . . . . . . . . Disabled0.000000 [mbar]
Low . . . . . . . . . . . . . . . . . . . . . . Disabled0.100000 [mbar]
High. . . . . . . . . . . . . . . . . . . . . . Enabled5.000000 [mbar]
High High . . . . . . . . . . . . . . . . . Disabled1.100000 [mbar]
Hysteresis . . . . . . . . . . . . . . . . . 1 [%]
Delay . . . . . . . . . . . . . . . . . . . . . 0 [s]
Filter State . . . . . . . . . . . . . . . . . Disabled
Type . . . . . . . . . . . . . . . . . . . . . Median
Depth . . . . . . . . . . . . . . . . . . . . 5
Central depth . . . . . . . . . . . . . . 3
Interferences
CO
2
Chlorinity/Salinity . . . . . . . . . . . All disabled19.000000 [g/l]
H
. . . . . . . . . . . . . . . . . . . . . . . Disabled0.100000 [bar]
2

9.1.2 Example of use

In this example we use:
One PC with a RS232 port.
One "RS-485<->RS232 converter"
Procedure:
1. Connect both RS-485 wires of the instrument to the "RS-485<->RS232 converter".
2. Connect the "RS-485<->RS232 converter" to the PC RS232 port using a standard cable
(RS232 DB9 straight cable).
On the PC:
1. Run "Hyperterminal" on the PC.
2. Configure the PC COM port used (e.g. COM2). Menu "File/Properties/Configure".
3. Configure the parameters "Baud rate", "Parity", "Nb of stop bits", "Byte size" (Menu
"File/Properties/Configure"). Use the same parameters for the instrument and the PC.
4. Configure the "Font = Courier 10" (Menu "View/Font").
5. Connect "Hyperterminal" (Menu "Call/Call").
6. Save the data received in the file of your choice (Menu "Transfer/Capture Text/Start").
On the instrument:
1. Use the menu "Communication/RS-485 Simple/Send files" and the button "All files".
When the transfer is finished, close the file with "Hyperterminal" (Menu "Transfer/Capture
Text/Stop"). Now, all the reports are saved in a text file on your PC.
or H
S . . . . . . . . . . . . . . . . All disabled
2
Communication Menu
69