7341 Calibration Bath
User's Guide
9.6.4
9.6.5
44
Pt=5
New soak time
Press "SET" to continue.
S
Save new setting

Program Function Mode

The next parameter is the program function or cycle mode. There are four pos-
sible modes which determine whether the program scans up (from set-point 1
to n) only or both up and down (from set-point n to 1), and also whether the
program stops after one cycle or repeats the cycle indefinitely. Table 3 below
shows the action of each of the four program mode settings.

Table 3 Program Mode Setting Actions

Function
Action
1
up-stop
2
up-down-stop
3
up-repeat
4
up-down-repeat
Program mode
Pf=1
Use the "UP" or "DOWN" buttons to change the mode.
Pf=4
New mode
Press "SET" to continue.
S
Enter program menu

Program Control

The final parameter in the program menu is the control parameter. Three op-
tions are available for controlling the ramp and soak program. The options are
to start the program from the beginning, (GO), continue the program from
where it was when it was stopped (Cont), or stop the program (OFF).
Pr=OFF
Program presently off
Use the "UP" or "DOWN" buttons to change the status.
Pr=Cont
Start cycle from beginning