7341 Calibration Bath
User's Guide
16
U
Increment display
New set-point value
C 30.00
Press SET to accept the new value and display the vernier value. The bath be-
gins heating or cooling to the new set-point.
S
Store new set-point, access vernier
Current vernier value
0.00000
Press "EXIT" and the bath temperature will be displayed again.
E
Return to the temperature display
24.73 C
Bath temperature display
The bath heats or cools until it reaches the new set-point temperature. Turn off
the cooling to reach and control at higher temperatures.
When setting the set-point temperature be careful not to exceed the temperature
limit of the bath fluid. The over-temperature cutout should be correctly set for
added safety. See Section 9.10, Cutout.
To obtain optimum control stability adjust the proportional band as discussed in
Section 9.9, Proportional Band.