Table 8. Calibration Adjustment Steps for 1577 Models with SN >32980000
Switch Position
(Function)
AC Volts
DC Volts
DC Millivolts
Ohms
Continuity
Diode Test
milliamps
Insulation
To avoid electrical shock hazard, remove all test leads from
the UUT before performing the following step.
Insulation
Notes:
[1]
Internal HV output calibration. Ensure no lead is inserted in Insulation (+) jack.
Input
Calibration Adjustment Step
Terminal
V/COM



















mA/COM




mA/COM

XWWarning
[1]
None

Insulation Multimeters
Calibration Adjustment Procedure
Input Value
600.0 mV, 60 Hz
6.000 V, 60 Hz
60.00 V, 60 Hz
600.0 V, 60 Hz
6.000 V, 0 Hz
60.00 V, 0 Hz
600.0 V, 0 Hz
600.0 mV, 0 Hz
60.00 mV, 0 Hz
600.0 Ω
6.000 kΩ
60.00 kΩ
600.0 kΩ
6.000 MΩ
0.000 Ω
50.0 MΩ
600.0 Ω
60.00 mA, 60 Hz
400.0 mA, 60 Hz
60.00 mA, 0 Hz
400.0 mA, 0 Hz
1.0 mA, 0 Hz
1.0 mA, 0 Hz
None (press BLUE to
start internal cal)
None (press BLUE to
start internal cal)