Available languages

Available languages

IMMUNITY test
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
IEC 60601 test level
<5 % UT (>95 % dip
in UT) for 0,5 cycle
40 % UT (60 % dip
in UT) for 5 cycles
70 % UT (30 % dip
in UT) for 25 cycles
<5 % UT (>95 % dip
in UT) for 5 s
3 A/m
Note: UT is the a.c. mains voltage prior to application of the test level.
Declaration – electromagnetic immunity
The Philips Avent breast pump is intended for use in the electromagnetic
environment specified below. The user of the Philips Avent breast pump
should assure that it is used in such an environment.
Compliance level
<5 % UT (>95 % dip
in UT) for 0,5 cycle
70 % UT (30 % dip
in UT) for 25 cycles
<5 % UT (>95 % dip
in UT) for 5 s
3 A/m
English
Electromagnetic
environment –
guidance
Mains power quality
should be that of a
typical residential or
hospital environment.
If the user of the Philips
Avent breast pump
requires continued
operation during power
mains interruptions, it
is recommended that
the Philips Avent breast
pump be powered
from an uninterruptible
power supply or a
battery.
Interruptible electric
supply can cause the
breast pump to turn
off. This is acceptable
as this will not lead to
unacceptable risks.
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical residential or
hospital environment.