Power Supply Wiring - Siemens SITRANS FUH1010PVDX Operating Instructions Manual

Ultrasonic flowmeters ip65 nema 7 compact precision volume 7me3601-3
Table of Contents

Advertisement

Section 2

2.2 .2 POWER SUPPLY WIRING

Using a 1/16" Hex key, loosen the Rear Housing Cover locking setscrew.
Unscrew the Rear Housing Cover and remove.
Locate power supply connector J8. Use the tables below and the part number of your unit to determine
the correct input power source. (See ID label on right side of housing.)
Power Supply Option Code
NOTE: Power Supply Module P/N label is located inside the front housing cover.
Pull the desired length of input power wires through
and into flowmeter case before wiring connector.
Wire input power connector P8 for AC or DC power
below.
Insert wires into wire entry holes and secure by tightening wire clamp screws (see figure
below).
NOTE: Power Supply connector wires should be stripped AWG 12 -18 stranded or
solid conductors.
Terminal 1010X-6SS2
2
1
P8
AC
NEU
HOT
DC
+
-
Stripped Wire
NOTE: The input power connector consists of a removable wire plug and PCB header located on the
1010X-8 I/O module. Two screws secure the wire plug to the header. The connector assembly is fully
polarized for fast and easy mating. The wire plug accepts stranded or solid conductors of AWG 12 - 18.
Typical Part Number Construction:
1010X-T2KNZPL2-S2----Safety Rating Option
Power Supply P/N
S
1010X-6SS2
ZN
1010X-6ZNS2
ZP
1010X-6ZPS2
Number
1
Hot
2
Neutral
3
Ground
3
1
2
GND
3
GND
Connector
Mounting
Screws
Input Power Connector (P8) Wiring
Root Meter Number
Transducer Support
Relay Support
Numeric Display
Power Supply Option
Language Option
User Supplied Power
90-250 VAC Single Phase
9-36 VDC Negative Ground
9-36 VDC Positive Ground
Flow Display Computer case conduit cable hole
depending on power supply provided
1010X-6ZNS2
1010X-6ZPS2
Positive
Positive (Gnd)
Negative (Gnd)
Negative
Ground
Ground
NOTE: Loosen Wire Clamp
Screw, insert stripped end
of wire and then retighten.
Wire
Clamp
Screws
2-2
CQO:1010PVXFM-3
as shown
Color Codes
Hot Neu GND
USA BK
WH GRN
EU
BR
BL Y/GRN

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sitrans fuh1010pvx

Table of Contents