Socket
16
Fig. 6: Installation in non-metallic vessel
1
Flange
2
Metal sheet
If possible, avoid sockets. Mount the sensor flush with the vessel top.
If this is not possible, use short sockets with small diameter.
Higher sockets or sockets with a bigger diameter can generally be
used. They can, however, increase the upper blocking distance (dead
band). Check if this is relevant for your measurement.
In such cases, always carry out a false signal suppression after instal-
lation. You can find further information under "Setup procedure".
Fig. 7: Mounting socket
When welding the socket, make sure that the socket is flush with the
vessel top.