Lifescan OneTouch SS Xpress Instructions For Use Manual page 99

Glucose hospital meter
Table of Contents

Advertisement

Analytical and System Accuracy Comparison: Glucose con-
centrations > 75 mg/dL
Site
Within
± 5%
Site #1
362/538
(67.3%)
Site #2
120/214
(56.1%)
Site #3
83/122
(68.0%)
Site #4
257/349
(73.6%)
Appendix
Table A-2 Arterial Specimens
Within
Within
± 10%
± 12.5%
497/538
515/538
(92.4%)
(95.7%)
195/214
208/214
(91.1%)
(97.2%)
112/122
115/122
(91.8%)
(94.3%)
330/349
340/349
(94.6%)
(97.4%)
Within
Within
± 15%
± 20%
523/538
532/538
(97.2%)
(98.9%)
214/214
214/214
(100%)
(100%)
117/122
120/122
(95.9%)
(98.4%)
346/349
348/349
(99.1%)
(99.7%)
Exceeds
± 20%
6/538
(1.1%)
0/214
(0.0%)
2/122
(1.6%)
1/349
(0.3%)