Download Print this page

Invacare Rea Dahlia electrical Service Manual page 2

Advertisement

Rea®Dahlia® electrical
1. Mount the attachment with the battery A on the frame.
Mounting the holder for the mains cable extension
1. Place the holder A on the marked area of the frame.
2. Attach the mains cable extension to the control box.
See section "Electrical schedule" for more
information about how to connect the cables.
WARNING!
Damage to the battery cable
– Make sure to fasten the cables in a way
that keeps the cables from being squeezed
or stretched.
– Check that the cables run freely when using
the tilt and/or recline function,
B
3. Re-mount the seat plate B.
4. Tighten the screws A.
3.1.1 Charging the battery
Damage to the battery
– The battery must be loaded 24 hours before
using the system the first time.
– Unplug the mains cable after loading before
using the wheelchair.
The battery charger has different charging cables in
order to adapt to different local electrical standards.
2
A
A
A
When the battery level is low (20 %), the system
beeps when it is in use.
3.2 Mounting the actuators
WARNING!
Risk of injury
The wheelchair may collapse
– Check that the locking washers are securely
locked.
– If the safety pin is used, make sure that the lock
shackle / loop is securely locked.
WARNING!
Safety risk
The wheelchair may collapse
– Remember to always reinsert and fasten the
safety pin when it has been removed.
– Check that the lock shackle / loop is securely
locked.
WARNING!
Electrical version
The backrest may come loose if the piston rod is
accidentally detached from its housing.
– Before attaching the piston to the backrest, turn
the piston rod C maximum clockwise.
– When aligning the holes, only turn the piston
rod C maximum a half turn counterclockwise.
1587500-A