Application With Stillpipe - Siemens SITRANS LR250 Operating Instructions Manual

Radar transmitters (ma/hart)
Hide thumbs Also See for SITRANS LR250:
Table of Contents

Advertisement

Parameter type
Quick Start Wiz-
ard parameters
Independent
parameters
Return to Measurement: press Mode
6.3.3

Application with stillpipe

Note
• Near Range (2.8.1.) (Blanking) will be set at the factory. Check the process connection
tag for specific values.
• Suitable pipe diameters are 40 mm (1.5") to 100 mm (4").
• The pipe diameter must be matched with the antenna size. Use the largest antenna size
that will fit the stillpipe/bypass pipe. See Dimensions (Page 195).
• See Mounting on a Stillpipe or Bypass Pipe (Page 31) for installation guidelines.
This application is to obtain a level measurement and corresponding 4 to 20 mA output
proportional to the oil level in a fuel storage vessel.
● Low Calibration Pt. is 5 m (16.5 ft) from the sensor reference point.
● High Calibration Pt. is 0.5 m (1.65 ft) from the sensor reference point.
● The stillpipe inside diameter is 50 mm (1.96").
● The maximum rate of filling or emptying is about 0.1 m (4")/min.
SITRANS LR250 (mA/HART)
Operating Instructions, 12/2015, A5E32220602-AD
Parameter No./Name
Introduction
Language
Material
Response Rate
Units
Operation
Low Calibration Point
High Calibration Point
Wizard Complete
Vessel Shape (2.7.1.1.)
Maximum Volume (2.7.1.2.)
Vessel Dimension A (2.7.1.3.)
Vessel Dimension L (2.7.1.4.)
LOE Timer (2.5.2.)
Material Level (2.5.1.)
to start normal operation.
6.3 Application examples
Options/
Function
Values
NEXT
Continue with Wizard.
NEXT
Continue with current
language.
LIQUID
MED
Medium =1 m/minute
M
meters
LEVEL
Level is reported as
Volume when a vessel
shape is selected.
3.5
3.5 m (11.48 ft)
0.5
0.5 m (1.64 ft)
FINISH
Transfers Quick Start
settings to device.
PARABOLIC
Defines vessel shape.
ENDS
8000
8000 liters
0.8
0.8 m (2.62 ft)
6
6 m (19.68 ft)
2
2 minutes
HI
Fail-safe level set to High
Commissioning
69

Advertisement

Table of Contents
loading

Table of Contents